EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A comparison of three‐film analysis software for stereotactic radiotherapy patient‐specific quality assurance

Athreya Buddhavarapu iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14203 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Athreya Buddhavarapu. Beteiligte Einrichtung: GenesisCare, Epping, Victoria, Australia. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

The aim of this study was to investigate the suitability of three radiochromic film analysis software for stereotactic radiotherapy patient‐specific quality assurance (PSQA): FilmQA Pro v5.0, SNC Patient v6.2, and eFilmQA v5.0.

Methods

Film calibration was conducted for each software followed by three sets of measurements. The first set assessed calibration accuracy by comparing measured and delivered doses at increments different from those used for calibration. The second set used each software to conduct PSQA through gamma analysis on 10 stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) patients. The third set utilized SNC Patient and eFilmQA to carry out gamma analysis on a collection of four digital test images, eliminating delivery and scanning uncertainties from impacting the analysis. Key supporting features within each software for conducting gamma analysis were identified.

Results

Overall, FilmQA Pro and eFilmQA were deemed comparable and favoured over SNC Patient due to the presence of key features such as triple‐channel dosimetry, auto‐optimization, and dose scaling. FilmQA Pro has a substantial user base and established reputation. eFilmQA, having been introduced more recently, serves as a viable alternative to FilmQA Pro, having been further refined for stereotactic radiotherapy PSQA.

Conclusion

This study investigated the suitability of three film analysis software (FilmQA Pro, eFilmQA, and SNC Patient) for stereotactic radiotherapy PSQA. Results from the investigation indicated that both FilmQA Pro and eFilmQA are comparably suitable and are preferred over SNC Patient. Both FilmQA Pro and eFilmQA are recommended for radiotherapy clinics.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryRadiotherapy, Intensity-ModulatedHumansFilm DosimetrySoftwareRadiotherapy DosageRadiometryRadiotherapy Planning, Computer-Assisted

Stichwörter der Autor:innen

medical physicsquality assuranceradiochromic filmstereotactic radiotherapy

Autor:innen und Institutionen

  • GenesisCare, Epping, Victoria, Australia.

ORCID-Profile: Athreya Buddhavarapu

Zitieren

APA

Buddhavarapu, A. (2024). A comparison of three‐film analysis software for stereotactic radiotherapy patient‐specific quality assurance. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14203

Vancouver

Buddhavarapu A. A comparison of three‐film analysis software for stereotactic radiotherapy patient‐specific quality assurance. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14203

BibTeX
@article{Buddhavarapu2024,
  author = {Buddhavarapu, Athreya},
  title = {A comparison of three‐film analysis software for stereotactic radiotherapy patient‐specific quality assurance},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14203}
}
Kostenlos starten →