EUVIMEDEUVIMED entdecken
Crossref

Site-specific performance and complexity-dependent risk assessment of patient-specific quality assurance using a 2D array detector in CyberKnife radiosurgery

Sandeep Singh, Supratik Sen, Dipesh, Abhay Kumar Singh, Manindra Bhushan, Soniya Pal, Benoy Kumar Singh, Raj Pal Singh, Anuj Vijay, Munish Gairola

Radiological Physics and Technology · 2026-12 · 19 (4) · 1684-1694

DOI: 10.1007/s12194-026-01136-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiological Physics and Technology, online erschienen am 05.09.2026. Verfasst von Sandeep Singh und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Amrita School of Medicine, Amrita Vishwavidyapeetham, Faridabad, Haryana, India. sandeep.singh82a@gmail.com. Hauptthemen laut MEDLINE: Radiosurgery, Quality Assurance, Health Care. Die Arbeit nennt 29 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To evaluate site-specific performance and complexity-dependent risk in patient-specific quality assurance for robotic radiosurgery using a two-dimensional array detector. A retrospective analysis was performed on 137 CyberKnife stereotactic treatment plans. PSQA was conducted using the PTW 1600 SRS 2D array detector. Gamma analysis was evaluated at multiple dose–distance criteria (3%/2 mm, 2%/2 mm, and 1%/1 mm). Delivery complexity was quantified using monitor units and robotic node count and combined into a Composite Complexity Index. Outlier analysis was performed using a clinically relevant action threshold of a gamma pass rate < 90% at 2%/2 mm. Across all sites, gamma pass rates exceeded 95% at 3%/2 mm, confirming robust baseline delivery accuracy. Progressive tightening of gamma criteria revealed pronounced site dependence, with lung and liver SBRT plans exhibiting the most significant degradation at 2%/2 mm and 1%/1 mm, respectively. Complexity effects were minimal at 3%/2 mm but became significant at 2%/2 mm and 1%/1 mm ( p < 0.01). Outlier analysis demonstrated low risk for brain and prostate treatments, intermediate risk for spine SBRT with complexity-dependent escalation, and consistently high outlier frequency for lung and liver SBRT independent of complexity. Array-based PSQA for robotic radiosurgery demonstrates high reliability under conventional gamma criteria while revealing strong site- and complexity-dependent behavior under stringent conditions. Anatomical site was the dominant determinant of QA variability and outlier risk. These findings support a risk-adaptive, site- and complexity-aware approach to PSQA interpretation rather than reliance on uniformly stringent acceptance thresholds.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryQuality Assurance, Health CareHumansRisk AssessmentRetrospective StudiesRadiotherapy Planning, Computer-Assisted

Stichwörter der Autor:innen

2D array detectorComposite complexity indexGamma index analysisQA risk stratificationRobotic radiosurgerypatient-specific quality assurance

Autor:innen und Institutionen

  • Department of Radiation Oncology, Amrita School of Medicine, Amrita Vishwavidyapeetham, Faridabad, Haryana, India. sandeep.singh82a@gmail.com.
  • Department of Physics, GLA University, Mathura, Uttar Pradesh, India. sandeep.singh82a@gmail.com.
  • Department of Radiation Oncology, Dr. B. Borooah Cancer Institute, Guwahati, Assam, India.
  • Department of Physics, GLA University, Mathura, Uttar Pradesh, India.
  • Department of Radiation Oncology, Andromeda Cancer Hospital, Kundli, Sonipat, Haryana, India.
  • Department of Radiation Oncology, Dharamshila Narayana Super speciality Hospital, New Delhi, India.

Zitieren

APA

Singh, S., Sen, S., Dipesh, Singh, A. K., Bhushan, M., Pal, S., Singh, B. K., Singh, R. P., Vijay, A., & Gairola, M. (2026). Site-specific performance and complexity-dependent risk assessment of patient-specific quality assurance using a 2D array detector in CyberKnife radiosurgery. Radiological Physics and Technology, 19(4), 1684-1694. https://doi.org/10.1007/s12194-026-01136-9

Vancouver

Singh S, Sen S, Dipesh, Singh AK, Bhushan M, Pal S, et al. Site-specific performance and complexity-dependent risk assessment of patient-specific quality assurance using a 2D array detector in CyberKnife radiosurgery. Radiological Physics and Technology. 2026;19(4):1684-1694. doi:10.1007/s12194-026-01136-9

BibTeX
@article{Singh2026,
  author = {Singh, Sandeep and Sen, Supratik and Dipesh and Singh, Abhay Kumar and Bhushan, Manindra and Pal, Soniya and Singh, Benoy Kumar and Singh, Raj Pal and Vijay, Anuj and Gairola, Munish},
  title = {Site-specific performance and complexity-dependent risk assessment of patient-specific quality assurance using a 2D array detector in CyberKnife radiosurgery},
  journal = {Radiological Physics and Technology},
  year = {2026},
  volume = {19},
  number = {4},
  pages = {1684--1694},
  doi = {10.1007/s12194-026-01136-9}
}
Kostenlos starten →