Improving the safety of radiotherapy treatment processes via incident‐driven FMEA feedback loops

Dominik Kornek iD, Michael Lotter, Juliane Szkitsak, Christopher Dürrbeck iD, Andre Karius iD, Oliver J Ott, Carolin Brandl, Christoph Bert iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14455 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Dominik Kornek und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Healthcare Failure Mode and Effect Analysis. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bavarian Ministry of Economic Affairs, Regional Development, and Energy.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Failure mode and effects analysis (FMEA) is a valuable tool for radiotherapy risk assessment, yet its outputs might be unreliable due to failures not being identified or due to a lack of accurate error rates.

Purpose

A novel incident reporting system (IRS) linked to an FMEA database was tested and evaluated. The study investigated whether the system was suitable for validating a previously performed analysis and whether it could provide accurate error rates to support the expert occurrence ratings of previously identified failure modes.

Methods

Twenty‐three pre‐identified failure modes of our external beam radiotherapy process, covering the process steps from patient admission to treatment delivery, were proffered on dedicated FMEA feedback and incident reporting terminals generated by the IRS. The clinical setting involved a computed tomography scanner, dosimetry, and five linacs. Incoming reports were used as basis to identify additional failure modes or confirm initial ones. The Kruskal–Wallis H test was applied to compare the risk priorities of the retrospective and prospective failure modes. Wald's sequential probability ratio test was used to investigate the correctness of the experts’ occurrence ratings by means of the number of incoming reports.

Results

Over a 15‐month period, 304 reports were submitted. There were 0.005 (confidence interval [CI], 0.0014–0.0082) reported incidents per imaging study and 0.0006 (CI, 0.0003–0.0009) reported incidents per treatment fraction. Sixteen additional failure modes could be identified, and their risk priorities did not differ from those of the initial failure modes ( p = 0.954). One failure mode occurrence rating could be increased, whereas the other 22 occurrence ratings could not be disproved.

Conclusions

Our approach is suitable for validating FMEAs and deducing additional failure modes on a continual basis. Accurate error rates can only be provided if a sufficient number of reports is available.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy DosageRadiotherapy Planning, Computer-AssistedHealthcare Failure Mode and Effect AnalysisHumansRisk ManagementRetrospective StudiesRadiotherapy, Intensity-ModulatedNeoplasmsRisk AssessmentFeedbackQuality Assurance, Health CarePatient SafetyProspective StudiesMedical ErrorsTomography, X-Ray Computed

Stichwörter der Autor:innen

error ratesfailure mode and effects analysis (FMEA)incident reporting system (IRS)radiotherapyrisk assessment

Autor:innen und Institutionen

  • Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Erlangen, Germany.

ORCID-Profile: Dominik Kornek, Christopher Dürrbeck, Andre Karius, Christoph Bert

Förderung

  • Bavarian Ministry of Economic Affairs, Regional Development, and Energy

Zitieren

APA

Kornek, D., Lotter, M., Szkitsak, J., Dürrbeck, C., Karius, A., Ott, O. J., Brandl, C., & Bert, C. (2024). Improving the safety of radiotherapy treatment processes via incident‐driven FMEA feedback loops. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14455

Vancouver

Kornek D, Lotter M, Szkitsak J, Dürrbeck C, Karius A, Ott OJ, et al. Improving the safety of radiotherapy treatment processes via incident‐driven FMEA feedback loops. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14455

BibTeX
@article{Kornek2024,
  author = {Kornek, Dominik and Lotter, Michael and Szkitsak, Juliane and Dürrbeck, Christopher and Karius, Andre and Ott, Oliver J and Brandl, Carolin and Bert, Christoph},
  title = {Improving the safety of radiotherapy treatment processes via incident‐driven FMEA feedback loops},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14455}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche