Pulse oximeter bench tests under different simulated skin tones
Medical & Biological Engineering & Computing · 2025-07 · 63 (7) · 1931-1942
DOI: 10.1007/s11517-024-03091-2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical & Biological Engineering & Computing, online erschienen am 24.04.2024. Verfasst von Suvvi K Narayana Swamy und 5 weiteren Autor:innen. Beteiligte Einrichtung: Optics and Photonics Research Group and Centre for Healthcare Technologies, University of Nottingham, University Park, Nottingham, UK. Hauptthemen laut MEDLINE: Oximetry, Skin Pigmentation. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medical Research Council, Medical Research Council & NHS@home programme, Engineering and Physical Sciences Research Council.
Abstract (Originalsprache)
Pulse oximeters’ (POs) varying performance based on skin tones has been highly publicised. Compared to arterial blood gas analysis, POs tend to overestimate oxygen saturation (SpO 2 ) values for people with darker skin (occult hypoxemia). The objective is to develop a test bench for assessing commercial home and hospital-based POs in controlled laboratory conditions. A laboratory simulator was used to mimic different SpO 2 values (~ 70 to 100%). Different neutral density and synthetic melanin filters were used to reproduce low signal and varying melanin attenuation levels. Six devices consisting of commercial home (Biolight, N = 13; ChoiceMMed, N = 18; MedLinket, N = 9) and hospital-based (Masimo Radical 7 with Neo L, N = 1; GE B450 Masimo SET with LNCS Neo L, N = 1; Nonin 9550 Onyx II™, N = 1) POs were reviewed and their response documented. Significant variations were observed in the recorded SpO 2 values among different POs when exposed to identical simulated signals. Differences were greatest for lower SpO 2 (< 80%) where empirical data is limited. All PO responses under low signal and melanin attenuation did not change across various simulated SpO 2 values. The bench tests do not provide conclusive evidence that melanin does not affect in vivo SpO 2 measurements. Research in the areas of instrument calibration, theory and design needs to be further developed. Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Optics and Photonics Research Group and Centre for Healthcare Technologies, University of Nottingham, University Park, Nottingham, UK.
- Clinical Engineering Department, Nottingham University Hospitals NHS Trust, Nottingham, UK.
- Optics and Photonics Research Group and Centre for Healthcare Technologies, University of Nottingham, University Park, Nottingham, UK. Steve.morgan@nottingham.ac.uk.
Förderung
- Medical Research Council
- Medical Research Council & NHS@home programme
- Engineering and Physical Sciences Research Council
Zitieren
APA
Swamy, S. K. N., He, C., Hayes-Gill, B. R., Clark, D. J., Green, S., & Morgan, S. P. (2025). Pulse oximeter bench tests under different simulated skin tones. Medical & Biological Engineering & Computing, 63(7), 1931-1942. https://doi.org/10.1007/s11517-024-03091-2
Vancouver
Swamy SKN, He C, Hayes-Gill BR, Clark DJ, Green S, Morgan SP. Pulse oximeter bench tests under different simulated skin tones. Medical & Biological Engineering & Computing. 2025;63(7):1931-1942. doi:10.1007/s11517-024-03091-2
BibTeX
@article{Swamy2025,
author = {Swamy, Suvvi K Narayana and He, Chenyang and Hayes-Gill, Barrie R and Clark, Daniel J and Green, Sarah and Morgan, Stephen P},
title = {Pulse oximeter bench tests under different simulated skin tones},
journal = {Medical & Biological Engineering & Computing},
year = {2025},
volume = {63},
number = {7},
pages = {1931--1942},
doi = {10.1007/s11517-024-03091-2}
}
EUVIMED