Accuracy of Rhythm Diagnostic Systems’ MultiSense® in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization

Charles Evrard, Amina El Attaoui iD, Cristina Pistea, Irina Enache, Mark Marriott, Louis Mayaud iD, Anne Charloux, Bernard Geny iD

Sensors · 2024

DOI: 10.3390/s25010127 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Charles Evrard und 7 weiteren Autor:innen. Beteiligte Einrichtung: Biomedicine Research Center of Strasbourg (CRBS), UR 3072, "Mitochondria, Oxidative Stress and Muscle Plasticity", Faculty of Medicine, University of…. Hauptthemen laut MEDLINE: Hypoxia, Oximetry, Oxygen, Oxygen Saturation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Rhytm diagnostic sysytem, ANR within the “Programme d’investissements d’avenir” France 2030.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The continuous monitoring of oxygen saturation (SpO2) and respiratory rates (RRs) are major clinical issues in many cardio-respiratory diseases and have been of tremendous importance during the COVID-19 pandemic. The early detection of hypoxemia was crucial since it precedes significant complications, and SpO2 follow-up allowed early hospital discharge in patients needing oxygen therapy. Nevertheless, fingertip devices showed some practical limitations. In this study, we investigated the reliability of the new Multisense® pulse oximetry system compared to a reference pulse oximeter (Vyntus CPX Pulse Oximeter) during hypoxia. In a population of sixteen healthy male subjects (mean age: 31.5 ± 7.0 years, BMI: 24.9 ± 3.6 kg/m², and 35% with darker skin tones), simultaneous SpO2 and RR measurements were collected over 12.4 h, during which FiO2 was progressively reduced from 21% to 10.5%. The average root mean square error (ARMS) of SpO2 for Multisense® placed on the back and chest was 2.94% and 2.98%, respectively, with permutation testing confirming a significant ARMS below 3.5% for both positions and no statistically significant difference in the ARMS between patch placements. Positive correlations and acceptable accuracy between devices were observed at both locations (r = 0.92, p < 0.001 and r = 0.90, p < 0.001 for back and chest placements, respectively). Bland–Altman analysis further indicated limits of agreement that support consistency across placements, with similar agreement levels noted across skin tones. Similar findings were obtained with the RR measurements. In conclusion, Multisense® demonstrated robust accuracy in measuring SpO2 and RRs during hypoxia in humans comparable to standard hospital-grade equipment. The effectiveness of the findings suggests that this wearable device is a valuable tool for the continuous monitoring of SpO2 and RRs, potentially enhancing patient safety and optimizing hospital resource allocation. Nevertheless, to overcome study limitations and allow generalized use, further work on a larger population sample, including more subjects with a high phototype and desaturation below 80%, would be useful.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HypoxiaOximetryOxygenOxygen SaturationRespiratory RateSkin PigmentationAdultHumansMaleCOVID-19Monitoring, PhysiologicReproducibility of Results

Stichwörter der Autor:innen

SpO2hypoxiapulse oximeterpulse oxygen saturationrespiratory ratesensorswearable device

Substanzen

Oxygen

Autor:innen und Institutionen

  • Biomedicine Research Center of Strasbourg (CRBS), UR 3072, "Mitochondria, Oxidative Stress and Muscle Plasticity", Faculty of Medicine, University of Strasbourg, 67000 Strasbourg, France.
  • Department of Physiology and Functional Explorations, University Hospital of Strasbourg, 67000 Strasbourg, France.
  • Rhythm Diagnostic Systems SAS, 67000 Strasbourg, France.
  • Institut Hospitalo-Universitaire, 67000 Strasbourg, France.

ORCID-Profile: Amina El Attaoui, Louis Mayaud, Bernard Geny

Förderung

  • Rhytm diagnostic sysytem
  • ANR within the “Programme d’investissements d’avenir” France 2030

Zitieren

APA

Evrard, C., El Attaoui, A., Pistea, C., Enache, I., Marriott, M., Mayaud, L., Charloux, A., & Geny, B. (2024). Accuracy of Rhythm Diagnostic Systems’ MultiSense® in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization. Sensors. https://doi.org/10.3390/s25010127

Vancouver

Evrard C, El Attaoui A, Pistea C, Enache I, Marriott M, Mayaud L, et al. Accuracy of Rhythm Diagnostic Systems’ MultiSense® in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization. Sensors. 2024. doi:10.3390/s25010127

BibTeX
@article{Evrard2024,
  author = {Evrard, Charles and El Attaoui, Amina and Pistea, Cristina and Enache, Irina and Marriott, Mark and Mayaud, Louis and Charloux, Anne and Geny, Bernard},
  title = {Accuracy of Rhythm Diagnostic Systems’ MultiSense® in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization},
  journal = {Sensors},
  year = {2024},
  doi = {10.3390/s25010127}
}
Kostenlos starten →