EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Effect of Skin Pigmentation and Finger Choice on Accuracy of Oxygen Saturation Measurement in an IoT-Based Pulse Oximeter

Shyqyri Haxha iD, Chike Nwibor, Mian Ali iD, Mohamed Sakel, Karen Saunders, Vladimir Dyo iD, Shakira Nabakooza

Sensors · 2024

DOI: 10.3390/s24113301 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Shyqyri Haxha und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronic Engineering, Royal Holloway, University of London, Egham TW20 0EX, UK. Hauptthemen laut MEDLINE: Oximetry, Skin Pigmentation, Fingers, Oxygen Saturation. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 63 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Pulse oximeters are widely used in hospitals and homes for measurement of blood oxygen saturation level (SpO2) and heart rate (HR). Concern has been raised regarding a possible bias in obtaining pulse oximeter measurements from different fingertips and the potential effect of skin pigmentation (white, brown, and dark). In this study, we obtained 600 SpO2 measurements from 20 volunteers using three UK NHS-approved commercial pulse oximeters alongside our custom-developed sensor, and used the Munsell colour system (5YR and 7.5YR cards) to classify the participants’ skin pigmentation into three distinct categories (white, brown, and dark). The statistical analysis using ANOVA post hoc tests (Bonferroni correction), a Bland–Altman plot, and a correlation test were then carried out to determine if there was clinical significance in measuring the SpO2 from different fingertips and to highlight if skin pigmentation affects the accuracy of SpO2 measurement. The results indicate that although the three commercial pulse oximeters had different means and standard deviations, these differences had no clinical significance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OximetrySkin PigmentationFingersOxygen SaturationHumansMaleAdultFemaleOxygenHeart RateYoung Adult

Stichwörter der Autor:innen

Internet of Thingsoptical sensorsphotoplethysmographypulse oximetry

Substanzen

Oxygen

Autor:innen und Institutionen

  • Department of Electronic Engineering, Royal Holloway, University of London, Egham TW20 0EX, UK.
  • East Kent Hospitals University NHS Foundation Trust, Canterbury CT2 7NT, UK.
  • Hertfordshire Partnership University NHS Foundation Trust, Hatfield, St. Albans AL3 5TQ, UK.

ORCID-Profile: Shyqyri Haxha, Mian Ali, Vladimir Dyo

Zitieren

APA

Haxha, S., Nwibor, C., Ali, M., Sakel, M., Saunders, K., Dyo, V., & Nabakooza, S. (2024). Effect of Skin Pigmentation and Finger Choice on Accuracy of Oxygen Saturation Measurement in an IoT-Based Pulse Oximeter. Sensors. https://doi.org/10.3390/s24113301

Vancouver

Haxha S, Nwibor C, Ali M, Sakel M, Saunders K, Dyo V, et al. Effect of Skin Pigmentation and Finger Choice on Accuracy of Oxygen Saturation Measurement in an IoT-Based Pulse Oximeter. Sensors. 2024. doi:10.3390/s24113301

BibTeX
@article{Haxha2024,
  author = {Haxha, Shyqyri and Nwibor, Chike and Ali, Mian and Sakel, Mohamed and Saunders, Karen and Dyo, Vladimir and Nabakooza, Shakira},
  title = {Effect of Skin Pigmentation and Finger Choice on Accuracy of Oxygen Saturation Measurement in an IoT-Based Pulse Oximeter},
  journal = {Sensors},
  year = {2024},
  doi = {10.3390/s24113301}
}
Kostenlos starten →