3D multi-phase balanced non-steady-state free precession acquisition for multi-parameter mapping

Riwaj Byanju, Gyula Kotek, Mika W Vogel, Stefan Klein, Juan A Hernandez-Tamames, Dirk H J Poot

Magnetic Resonance Materials in Physics, Biology and Medicine · 2025

DOI: 10.1007/s10334-025-01262-2 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2025. Verfasst von Riwaj Byanju und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, The Netherlands. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Magnetic Resonance Imaging. Die Arbeit nennt 65 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: GE Healthcare.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study presents the 3D MP-b-nSSFP sequence for multi-parametric mapping.

Methods

We evaluate several aspects of the 3D implementation, like the type of RF pulse (selective/non-selective), the readout duration, the undersampling pattern, and the acceleration factor. We use undersampled scans with subspace-constrained reconstruction and extended spiral readouts to achieve clinically acceptable scan times. The repeatability and accuracy of the $$T_1$$ T 1 and $$T_2$$ T 2 maps are compared with a reference technique in phantom and three volunteer scans.

Results

Compared with selective refocusing pulses, we observe lower bias with non-selective pulses, despite modeling the spatially varying effect of the pulses in the fitting process. $$T_1$$ T 1 and $$T_2$$ T 2 maps obtained from phantom scans were comparable to the nominal values and those from reference scans. $$T_1$$ T 1 values in vivo were underestimated compared to the reference scan. The maps with an acquisition matrix of $$256 \times 256 \times 44$$ 256 × 256 × 44 and resolution $$1 \times 1 \times 3$$ 1 × 1 × 3 $$\text {mm}^3$$ mm 3 were acquired in 11 min.

Conclusion

We show that 3D MP-b-nSSFP can be used for multi-parameter mapping within clinically acceptable scan time. Phantom scans show results in good agreement with reference scan results. However, the in vivo scan underestimated $$T_1$$ T 1 .

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Imaging, Three-DimensionalMagnetic Resonance ImagingHumansPhantoms, ImagingAlgorithmsReproducibility of ResultsBrainImage Processing, Computer-AssistedImage Interpretation, Computer-AssistedAdult

Stichwörter der Autor:innen

Extended readoutSSFPSubspace-constrainedTransient imagingqMRI

Autor:innen und Institutionen

  • Department of Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, The Netherlands.
  • GE Healthcare, Hoevelaken, The Netherlands.
  • Department of Radiology and Nuclear Medicine, Erasmus MC, Rotterdam, The Netherlands. d.poot@erasmusmc.nl.

Förderung

  • GE Healthcare

Zitieren

APA

Byanju, R., Kotek, G., Vogel, M. W., Klein, S., Hernandez-Tamames, J. A., & Poot, D. H. J. (2025). 3D multi-phase balanced non-steady-state free precession acquisition for multi-parameter mapping. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-025-01262-2

Vancouver

Byanju R, Kotek G, Vogel MW, Klein S, Hernandez-Tamames JA, Poot DHJ. 3D multi-phase balanced non-steady-state free precession acquisition for multi-parameter mapping. Magnetic Resonance Materials in Physics, Biology and Medicine. 2025. doi:10.1007/s10334-025-01262-2

BibTeX
@article{Byanju2025,
  author = {Byanju, Riwaj and Kotek, Gyula and Vogel, Mika W and Klein, Stefan and Hernandez-Tamames, Juan A and Poot, Dirk H J},
  title = {3D multi-phase balanced non-steady-state free precession acquisition for multi-parameter mapping},
  journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
  year = {2025},
  doi = {10.1007/s10334-025-01262-2}
}
Kostenlos starten →