Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 2): T1 and T2 standardization in the ISMRM/NIST MR phantom

Siria Pasini, Steffen Ringgaard, Tau Vendelboe, Leyre Garcia-Ruiz, Anika Strittmatter, Giulia Villa, Anish Raj, Rebeca Echeverria-Chasco, Michela Bozzetto, Paolo Brambilla, Malene Aastrup, Esben S S Hansen und 7 weitere

Magnetic Resonance Materials in Physics, Biology and Medicine · 17.05.2025 · 38 (3) · 611-627

DOI: 10.1007/s10334-025-01260-4 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, online erschienen am 17.05.2025. Verfasst von Siria Pasini und 18 weiteren Autor:innen. Beteiligte Einrichtung: Bioengineering Department, Istituto di Ricerche Farmacologiche Mario Negri IRCCS, Via Camozzi 3, 24020, Ranica, BG, Italy. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Ministero della Salute, Gobierno de Navarra, Bundesministerium für Bildung und Forschung.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: To assess multi-site and multi-vendor accuracy, and intra- and inter-scanner variability of T1 and T2 measurements using the ISMRM/NIST System MRI phantom at room temperature. MATERIALS AND METHODS: T1 and T2 measurements were acquired using standardized NIST protocols on 13 scanners (1.5 T and 3 T) from 3 vendors at 7 sites and compared with reference values at room temperature. Pearson's correlation (r) and accuracy error were used for comparison with reference values, while inter-scanner agreement was assessed using the coefficient of variation (CV%). Short-term reproducibility was evaluated using Bland-Altman plots and precision error. Generalized linear mixed models and post hoc tests (α = 0.05) were adopted to compare accuracy and precision across field strengths, vendors, and scanners. T2 measurements were corrected with StimFit toolbox for stimulated echo compensation. RESULTS: T1 and T2 measurements had excellent correlation with reference values at both field strengths. Stimfit significantly improved T2 accuracy in the renal range for 9 of 13 scanners. Short-term reproducibility (limits of agreement < 10%) and inter-scanner agreement were good (median CV < 7%) for both T1 and T2 values. Inter-scanner CV was < 5% in the renal range for both parameters. DISCUSSION: These findings support the need of scanner evaluation processes to ensure reliable T1-T2 measurements in multi-center MRI studies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingReproducibility of ResultsHumansReference StandardsImage Processing, Computer-AssistedReference ValuesTemperatureLinear Models

Stichwörter der Autor:innen

MRIMulti-sitePhantomRelaxation timesStandardization

Autor:innen und Institutionen

  • Bioengineering Department, Istituto di Ricerche Farmacologiche Mario Negri IRCCS, Via Camozzi 3, 24020, Ranica, BG, Italy.
  • The MR Research Centre, Aarhus University, Aarhus, Denmark.
  • Department of Radiology, Clínica Universidad de Navarra, Pamplona, Spain.
  • Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
  • Mannheim Institute for Intelligent Systems in Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
  • Radiology Unit, ASST Papa Giovanni XXIII, Bergamo, Italy.

Förderung

  • Ministero della Salute
  • Gobierno de Navarra
  • Bundesministerium für Bildung und Forschung
  • Innovationsfonden
  • “Aiuti per la Ricerca sulle Malattie Rare” (A.R.M.R) Foundation

Zitieren

APA

Pasini, S., Ringgaard, S., Vendelboe, T., Garcia-Ruiz, L., Strittmatter, A., Villa, G., Raj, A., Echeverria-Chasco, R., Bozzetto, M., Brambilla, P., Aastrup, M., Hansen, E. S. S., Pierotti, L., Renzulli, M., Francis, S. T., Zöllner, F. G., Laustsen, C., Fernandez-Seara, M. A., & Caroli, A. (2025). Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 2): T1 and T2 standardization in the ISMRM/NIST MR phantom. Magnetic Resonance Materials in Physics, Biology and Medicine, 38(3), 611-627. https://doi.org/10.1007/s10334-025-01260-4

Vancouver

Pasini S, Ringgaard S, Vendelboe T, Garcia-Ruiz L, Strittmatter A, Villa G, et al. Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 2): T1 and T2 standardization in the ISMRM/NIST MR phantom. Magnetic Resonance Materials in Physics, Biology and Medicine. 2025;38(3):611-627. doi:10.1007/s10334-025-01260-4

BibTeX
@article{Pasini2025,
  author = {Pasini, Siria and Ringgaard, Steffen and Vendelboe, Tau and Garcia-Ruiz, Leyre and Strittmatter, Anika and Villa, Giulia and Raj, Anish and Echeverria-Chasco, Rebeca and Bozzetto, Michela and Brambilla, Paolo and Aastrup, Malene and Hansen, Esben S S and Pierotti, Luisa and Renzulli, Matteo and Francis, Susan T and Zöllner, Frank G and Laustsen, Christoffer and Fernandez-Seara, Maria A and Caroli, Anna},
  title = {Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 2): T1 and T2 standardization in the ISMRM/NIST MR phantom},
  journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
  year = {2025},
  volume = {38},
  number = {3},
  pages = {611--627},
  doi = {10.1007/s10334-025-01260-4}
}
Kostenlos starten →