Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 1): apparent diffusion coefficient standardization in a diffusion MRI phantom

Siria Pasini, Steffen Ringgaard, Tau Vendelboe, Leyre Garcia-Ruiz, Anika Strittmatter, Giulia Villa, Anish Raj, Rebeca Echeverria-Chasco, Michela Bozzetto, Paolo Brambilla, Malene Aastrup, Esben S S Hansen und 7 weitere

Magnetic Resonance Materials in Physics, Biology and Medicine · 2025

DOI: 10.1007/s10334-025-01256-0 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2025. Verfasst von Siria Pasini und 18 weiteren Autor:innen. Beteiligte Einrichtung: Bioengineering Department, Istituto di Ricerche Farmacologiche Mario Negri IRCCS, Via Camozzi 3, 24020, Ranica, BG, Italy. Hauptthemen laut MEDLINE: Phantoms, Imaging, Diffusion Magnetic Resonance Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Ministero della Salute, Gobierno de Navarra, Bundesministerium für Bildung und Forschung.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: To validate multi-site and multi-vendor ADC measurements using the QIBA/NIST diffusion MRI phantom at room temperature. MATERIALS AND METHODS: ADC measurements were performed on 12 scanners (evenly split between 1.5 and 3 T) from three vendors at five sites and compared with reference values at room temperature. We adopted Pearson's correlation (r) and accuracy error for comparison with reference values; within scanner coefficient of variation (CVintra%) for intra-session repeatability and inter-scanner for agreement (CVinter%); Bland-Altman plots and precision error for short-term reproducibility; generalized linear mixed models and post-hoc tests ( α =0.05) to compare accuracy, repeatability and precision across field strengths, vendors, and scanners. RESULTS: Temperature adjusted ADCs were well correlated with NIST reference values (r ≥ 0.997 for 1.5 T, r ≥ 0.996 for 3 T). Median accuracy error was lower than 5% for all scanners. In the renal physiologic range (ADC > 0.83 × 10-3 mm2/s), accuracy error was < 10% and CVintra < 2%. Across all scanners, good short-term reproducibility with limits of agreement < 10% and excellent agreement (median CVinter < 2%) were found. DISCUSSION: Despite using abdominal receive coils and room temperature measurements, all quantitative parameters were within literature findings. High accuracy, repeatability and precision within the renal physiologic range support the feasibility of scanner evaluation using QIBA standardization process for diffusion measurements in renal studies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingDiffusion Magnetic Resonance ImagingReproducibility of ResultsHumansTemperatureReference StandardsReference ValuesImage Processing, Computer-AssistedLinear Models

Stichwörter der Autor:innen

Apparent diffusion coefficientMRIMulti-sitePhantomStandardization

Autor:innen und Institutionen

  • Bioengineering Department, Istituto di Ricerche Farmacologiche Mario Negri IRCCS, Via Camozzi 3, 24020, Ranica, BG, Italy.
  • The MR Research Centre, Aarhus University, Aarhus, Denmark.
  • Department of Radiology, Clínica Universidad de Navarra, Pamplona, Spain.
  • Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
  • Mannheim Institute for Intelligent Systems in Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
  • Radiology Unit, ASST Papa Giovanni XXIII, Bergamo, Italy.

Förderung

  • Ministero della Salute
  • Gobierno de Navarra
  • Bundesministerium für Bildung und Forschung
  • Innovationsfonden
  • “Aiuti per la Ricerca sulle Malattie Rare” (A.R.M.R) Foundation

Zitieren

APA

Pasini, S., Ringgaard, S., Vendelboe, T., Garcia-Ruiz, L., Strittmatter, A., Villa, G., Raj, A., Echeverria-Chasco, R., Bozzetto, M., Brambilla, P., Aastrup, M., Hansen, E. S. S., Pierotti, L., Renzulli, M., Francis, S. T., Zoellner, F. G., Laustsen, C., Fernandez-Seara, M. A., & Caroli, A. (2025). Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 1): apparent diffusion coefficient standardization in a diffusion MRI phantom. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-025-01256-0

Vancouver

Pasini S, Ringgaard S, Vendelboe T, Garcia-Ruiz L, Strittmatter A, Villa G, et al. Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 1): apparent diffusion coefficient standardization in a diffusion MRI phantom. Magnetic Resonance Materials in Physics, Biology and Medicine. 2025. doi:10.1007/s10334-025-01256-0

BibTeX
@article{Pasini2025,
  author = {Pasini, Siria and Ringgaard, Steffen and Vendelboe, Tau and Garcia-Ruiz, Leyre and Strittmatter, Anika and Villa, Giulia and Raj, Anish and Echeverria-Chasco, Rebeca and Bozzetto, Michela and Brambilla, Paolo and Aastrup, Malene and Hansen, Esben S S and Pierotti, Luisa and Renzulli, Matteo and Francis, Susan T and Zoellner, Frank G and Laustsen, Christoffer and Fernandez-Seara, Maria A and Caroli, Anna},
  title = {Multi-center and multi-vendor evaluation study across 1.5 T and 3 T scanners (part 1): apparent diffusion coefficient standardization in a diffusion MRI phantom},
  journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
  year = {2025},
  doi = {10.1007/s10334-025-01256-0}
}
Kostenlos starten →