Assessing Measurement Repeatability of a Novel Anisotropic Phantom for Advanced Diffusion MRI Models
Magnetic Resonance in Medicine · 2026-07 · 96 (1) · 339-348
DOI: 10.1002/mrm.70330 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 06.03.2026. Verfasst von Lauren Stephens und 3 weiteren Autor:innen. Beteiligte Einrichtung: McMaster School of Biomedical Engineering, McMaster University, Hamilton, Ontario, Canada. Hauptthemen laut MEDLINE: Phantoms, Imaging, Diffusion Magnetic Resonance Imaging, White Matter, Image Processing, Computer-Assisted. Die Arbeit nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Southern Ontario Pharmaceutical and Health Innovation Ecosystem (SOPHIE).
Abstract (Originalsprache)
Purpose
Diffusion MRI is widely used to characterize tissue microstructure, but standardization remains challenging, particularly for advanced models or regions with crossing fibers. Phantoms provide controlled environments to assess measurement repeatability independent of biological variability. This study evaluated the repeatability of higher‐order diffusion tensor metrics using a novel anisotropic diffusion phantom designed to mimic white matter tract geometry.
Methods
The phantom, containing linear, crossing (30°, 45°, 90°), and bifurcating synthetic fiber bundles, was scanned seven times using a GE Healthcare 3.0 T MRI system. Four acquisition protocols were evaluated: 30‐direction DTI ( b = 1000s/mm 2 ), 60 and 90‐direction High Angular Resolution Diffusion Imaging (HARDI; b = 1300s/mm 2 ), and 30‐direction Diffusion Kurtosis Imaging (DKI; b = 250, 500, 750, 1000, 1500, 2000, 2500, 3000 s/mm 2 ). Repeatability was quantified using coefficient of variation (CoV) and intraclass correlation coefficient (ICC) for scalar diffusion metrics across six regions of interest. Fiber orientation distribution functions (fODFs) were analyzed to assess crossing fiber resolution accuracy.
Results
DTI‐derived metrics demonstrated excellent repeatability, with fractional anisotropy (FA) CoV < 10% and mean, axial, and radial diffusivities < 3%. DKI‐derived metrics exhibited greater variability, though kurtosis FA remained stable (CoV ∼7%). Generalized FA showed improved reliability with increased angular resolution (ICC = 0.8445 for 90‐direction HARDI). fODFs accurately resolved crossing fibers at 90° (RMSE = 3.49°) and 45° (RMSE = 8.92°) but failed at 30° separation.
Conclusion
The phantom provides reliable repeatability for standard DTI metrics and demonstrates utility for quality assurance of advanced diffusion models with high angular resolution protocols.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- McMaster School of Biomedical Engineering, McMaster University, Hamilton, Ontario, Canada.
- Imaging Research Centre, St. Joseph's Healthcare, Hamilton, Ontario, Canada.
- Department of Medical Imaging, McMaster University, Hamilton, Ontario, Canada.
- PreOperative Performance, Toronto, Ontario, Canada.
- Department of Electrical and Computer Engineering, McMaster University, Hamilton, Ontario, Canada.
ORCID-Profile: Lauren Stephens, Sofia Chavez, Fergal Kerins, Michael D Noseworthy
Förderung
- Southern Ontario Pharmaceutical and Health Innovation Ecosystem (SOPHIE)
Zitieren
APA
Stephens, L., Chavez, S., Kerins, F., & Noseworthy, M. D. (2026). Assessing Measurement Repeatability of a Novel Anisotropic Phantom for Advanced Diffusion MRI Models. Magnetic Resonance in Medicine, 96(1), 339-348. https://doi.org/10.1002/mrm.70330
Vancouver
Stephens L, Chavez S, Kerins F, Noseworthy MD. Assessing Measurement Repeatability of a Novel Anisotropic Phantom for Advanced Diffusion MRI Models. Magnetic Resonance in Medicine. 2026;96(1):339-348. doi:10.1002/mrm.70330
BibTeX
@article{Stephens2026,
author = {Stephens, Lauren and Chavez, Sofia and Kerins, Fergal and Noseworthy, Michael D},
title = {Assessing Measurement Repeatability of a Novel Anisotropic Phantom for Advanced Diffusion MRI Models},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {1},
pages = {339--348},
doi = {10.1002/mrm.70330}
}