Fast bias-corrected conductivity mapping using stimulated echoes
Magnetic Resonance Materials in Physics, Biology and Medicine · 2024
DOI: 10.1007/s10334-024-01194-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2024. Verfasst von Santhosh Iyyakkunnel und 2 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiological Physics, Department of Radiology , University Hospital Basel, Basel, Switzerland. santhosh.iyyakkunnel@unibas.ch. Hauptthemen laut MEDLINE: Phantoms, Imaging, Electric Conductivity, Brain, Algorithms. Die Arbeit nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung, University of Basel.
Abstract (Originalsprache)
Objective
To demonstrate the potential of a double angle stimulated echo (DA-STE) method for fast and accurate “full” homogeneous Helmholtz-based electrical properties tomography using a simultaneous $${B}_{1}^{+}$$ B 1 + magnitude and transceive phase measurement.
Methods
The combination of a spin and stimulated echo can be used to yield an estimate of both $${B}_{1}^{+}$$ B 1 + magnitude and transceive phase and thus provides the means for “full" EPT reconstruction. An interleaved 2D acquisition scheme is used for rapid acquisition. The method was validated in a saline phantom and compared to a double angle method based on two single gradient echo acquisitions (GRE-DAM). The method was evaluated in the brain of a healthy volunteer.
Results
The $${B}_{1}^{+}$$ B 1 + magnitude obtained with DA-STE showed excellent agreement with the GRE-DAM method. Conductivity values based on the “full” EPT reconstruction also agreed well with the expectations in the saline phantom. In the brain, the method delivered conductivity values close to literature values.
Discussion
The method allows the use of the “full” Helmholtz-based EPT reconstruction without the need for additional measurements. As a result, quantitative conductivity values are improved compared to phase-based EPT reconstructions. DA-STE is a fast complex- $${B}_{1}^{+}$$ B 1 + mapping technique that could render EPT clinically relevant at 3 T.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Radiological Physics, Department of Radiology , University Hospital Basel, Basel, Switzerland. santhosh.iyyakkunnel@unibas.ch.
- Department of Biomedical Engineering, University of Basel, Basel, Switzerland. santhosh.iyyakkunnel@unibas.ch.
- Division of Radiological Physics, Department of Radiology , University Hospital Basel, Basel, Switzerland.
- Department of Biomedical Engineering, University of Basel, Basel, Switzerland.
- Translational Imaging in Neurology (ThINk) Basel, Department of Biomedical Engineering, Faculty of Medicine, University Hospital Basel and University of Basel, Basel, Switzerland.
- Neurologic Clinic and Policlinic, MS Center and Research Center for Clinical Neuroimmunology and Neuroscience Basel (RC2NB), University Hospital Basel and University of Basel, Basel, Switzerland.
Förderung
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
- University of Basel
Zitieren
APA
Iyyakkunnel, S., Weigel, M., & Bieri, O. (2024). Fast bias-corrected conductivity mapping using stimulated echoes. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-024-01194-3
Vancouver
Iyyakkunnel S, Weigel M, Bieri O. Fast bias-corrected conductivity mapping using stimulated echoes. Magnetic Resonance Materials in Physics, Biology and Medicine. 2024. doi:10.1007/s10334-024-01194-3
BibTeX
@article{Iyyakkunnel2024,
author = {Iyyakkunnel, Santhosh and Weigel, Matthias and Bieri, Oliver},
title = {Fast bias-corrected conductivity mapping using stimulated echoes},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2024},
doi = {10.1007/s10334-024-01194-3}
}
EUVIMED