Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise

Anne M Mulholland, Hayley V MacDonald, Elroy J Aguiar, Jonathan E Wingo

European Journal of Applied Physiology · 2026-02 · 126 (2) · 1057-1066

DOI: 10.1007/s00421-025-05977-x ↗

At a glance

Journal article in European Journal of Applied Physiology, published online on 2025-09-18. Written by Anne M Mulholland and 3 co-authors. Institution: Department of Exercise Science, Mercer University, Macon, GA, USA. mulholland_a@mercer.edu. Major MEDLINE topics: Photoplethysmography, Skin Pigmentation, Heart Rate, Exercise. Cited 12 times so far (Crossref); 48 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Introduction

Wearable devices often use photoplethysmography to estimate heart rate (HR) by measuring light reflected from the cutaneous vascular bed. Notably, higher melanin content of the epidermis may reduce the amount of light transmitted through the skin. Previous studies examining the impact of skin tone on photoplethysmographic HR measurement accuracy have produced mixed results; however, none measured epidermal melanin content.

Purpose

To determine whether objectively measured skin pigmentation influences the accuracy of photoplethysmographic HR measurement during rest, exercise, and recovery.

Methods

Skin pigmentation was quantified from colorimeter measures using Individual Typology Angle (ITA°), a strong correlate of epidermal melanin, in 28 healthy adults (White, n = 16; Black, n = 10; Middle Eastern, n = 1; Persian, n = 1). Chest-worn HR (criterion; Polar H10) and photoplethysmographic HR from three devices (Apple Watch Series 8, Garmin vivosmart 5, SlateSafety BAND V2) were recorded continuously during rest, cycling, and recovery; HR data were averaged into 30-s epochs for analysis. A linear mixed-effects model determined whether ITA° influenced mean absolute error of HR (MAE HR ) for each device.

Results

ITA° predicted MAE HR for SlateSafety (β = − 0.011, P = 0.001), but not for Apple ( P = 0.62) or Garmin ( P = 0.29). Missing data were disproportionately attributed to participants with dark skin (ITA° < 10°; 36%) for Apple (50%) and SlateSafety (85%) devices; outliers were disproportionately observed among participants with dark skin for all devices (56%–62%).

Conclusion

Data quality was negatively impacted by darker skin pigmentation for all tested devices, but skin pigmentation (ITA°) only increased photoplethysmographic HR measurement error to a small degree (~ 1 bpm) for the SlateSafety device.

Subjects

MeSH terms (MEDLINE)

PhotoplethysmographySkin PigmentationHeart RateExerciseHumansAdultMaleFemaleYoung AdultMelanins

Author keywords

Physiological monitoringSkin colorWearable deviceWearable technology

Substances

Melanins

Authors and institutions

  • Department of Exercise Science, Mercer University, Macon, GA, USA. mulholland_a@mercer.edu.
  • Department of Kinesiology, The University of Alabama, Tuscaloosa, AL, USA. mulholland_a@mercer.edu.
  • Department of Kinesiology, The University of Alabama, Tuscaloosa, AL, USA.

Cite

APA

Mulholland, A. M., MacDonald, H. V., Aguiar, E. J., & Wingo, J. E. (2026). Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise. European Journal of Applied Physiology, 126(2), 1057-1066. https://doi.org/10.1007/s00421-025-05977-x

Vancouver

Mulholland AM, MacDonald HV, Aguiar EJ, Wingo JE. Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise. European Journal of Applied Physiology. 2026;126(2):1057-1066. doi:10.1007/s00421-025-05977-x

BibTeX
@article{Mulholland2026,
  author = {Mulholland, Anne M and MacDonald, Hayley V and Aguiar, Elroy J and Wingo, Jonathan E},
  title = {Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise},
  journal = {European Journal of Applied Physiology},
  year = {2026},
  volume = {126},
  number = {2},
  pages = {1057--1066},
  doi = {10.1007/s00421-025-05977-x}
}
Start for free →