Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise

Anne M Mulholland, Hayley V MacDonald, Elroy J Aguiar, Jonathan E Wingo

European Journal of Applied Physiology · 2026-02 · 126 (2) · 1057-1066

DOI: 10.1007/s00421-025-05977-x ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Journal of Applied Physiology, online erschienen am 18.09.2025. Verfasst von Anne M Mulholland und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Exercise Science, Mercer University, Macon, GA, USA. mulholland_a@mercer.edu. Hauptthemen laut MEDLINE: Photoplethysmography, Skin Pigmentation, Heart Rate, Exercise. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Wearable devices often use photoplethysmography to estimate heart rate (HR) by measuring light reflected from the cutaneous vascular bed. Notably, higher melanin content of the epidermis may reduce the amount of light transmitted through the skin. Previous studies examining the impact of skin tone on photoplethysmographic HR measurement accuracy have produced mixed results; however, none measured epidermal melanin content.

Purpose

To determine whether objectively measured skin pigmentation influences the accuracy of photoplethysmographic HR measurement during rest, exercise, and recovery.

Methods

Skin pigmentation was quantified from colorimeter measures using Individual Typology Angle (ITA°), a strong correlate of epidermal melanin, in 28 healthy adults (White, n = 16; Black, n = 10; Middle Eastern, n = 1; Persian, n = 1). Chest-worn HR (criterion; Polar H10) and photoplethysmographic HR from three devices (Apple Watch Series 8, Garmin vivosmart 5, SlateSafety BAND V2) were recorded continuously during rest, cycling, and recovery; HR data were averaged into 30-s epochs for analysis. A linear mixed-effects model determined whether ITA° influenced mean absolute error of HR (MAE HR ) for each device.

Results

ITA° predicted MAE HR for SlateSafety (β = − 0.011, P = 0.001), but not for Apple ( P = 0.62) or Garmin ( P = 0.29). Missing data were disproportionately attributed to participants with dark skin (ITA° < 10°; 36%) for Apple (50%) and SlateSafety (85%) devices; outliers were disproportionately observed among participants with dark skin for all devices (56%–62%).

Conclusion

Data quality was negatively impacted by darker skin pigmentation for all tested devices, but skin pigmentation (ITA°) only increased photoplethysmographic HR measurement error to a small degree (~ 1 bpm) for the SlateSafety device.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhotoplethysmographySkin PigmentationHeart RateExerciseHumansAdultMaleFemaleYoung AdultMelanins

Stichwörter der Autor:innen

Physiological monitoringSkin colorWearable deviceWearable technology

Substanzen

Melanins

Autor:innen und Institutionen

  • Department of Exercise Science, Mercer University, Macon, GA, USA. mulholland_a@mercer.edu.
  • Department of Kinesiology, The University of Alabama, Tuscaloosa, AL, USA. mulholland_a@mercer.edu.
  • Department of Kinesiology, The University of Alabama, Tuscaloosa, AL, USA.

Zitieren

APA

Mulholland, A. M., MacDonald, H. V., Aguiar, E. J., & Wingo, J. E. (2026). Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise. European Journal of Applied Physiology, 126(2), 1057-1066. https://doi.org/10.1007/s00421-025-05977-x

Vancouver

Mulholland AM, MacDonald HV, Aguiar EJ, Wingo JE. Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise. European Journal of Applied Physiology. 2026;126(2):1057-1066. doi:10.1007/s00421-025-05977-x

BibTeX
@article{Mulholland2026,
  author = {Mulholland, Anne M and MacDonald, Hayley V and Aguiar, Elroy J and Wingo, Jonathan E},
  title = {Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise},
  journal = {European Journal of Applied Physiology},
  year = {2026},
  volume = {126},
  number = {2},
  pages = {1057--1066},
  doi = {10.1007/s00421-025-05977-x}
}
Kostenlos starten →