Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise
European Journal of Applied Physiology · 2026-02 · 126 (2) · 1057-1066
DOI: 10.1007/s00421-025-05977-x ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Applied Physiology, online erschienen am 18.09.2025. Verfasst von Anne M Mulholland und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Exercise Science, Mercer University, Macon, GA, USA. mulholland_a@mercer.edu. Hauptthemen laut MEDLINE: Photoplethysmography, Skin Pigmentation, Heart Rate, Exercise. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Wearable devices often use photoplethysmography to estimate heart rate (HR) by measuring light reflected from the cutaneous vascular bed. Notably, higher melanin content of the epidermis may reduce the amount of light transmitted through the skin. Previous studies examining the impact of skin tone on photoplethysmographic HR measurement accuracy have produced mixed results; however, none measured epidermal melanin content.
Purpose
To determine whether objectively measured skin pigmentation influences the accuracy of photoplethysmographic HR measurement during rest, exercise, and recovery.
Methods
Skin pigmentation was quantified from colorimeter measures using Individual Typology Angle (ITA°), a strong correlate of epidermal melanin, in 28 healthy adults (White, n = 16; Black, n = 10; Middle Eastern, n = 1; Persian, n = 1). Chest-worn HR (criterion; Polar H10) and photoplethysmographic HR from three devices (Apple Watch Series 8, Garmin vivosmart 5, SlateSafety BAND V2) were recorded continuously during rest, cycling, and recovery; HR data were averaged into 30-s epochs for analysis. A linear mixed-effects model determined whether ITA° influenced mean absolute error of HR (MAE HR ) for each device.
Results
ITA° predicted MAE HR for SlateSafety (β = − 0.011, P = 0.001), but not for Apple ( P = 0.62) or Garmin ( P = 0.29). Missing data were disproportionately attributed to participants with dark skin (ITA° < 10°; 36%) for Apple (50%) and SlateSafety (85%) devices; outliers were disproportionately observed among participants with dark skin for all devices (56%–62%).
Conclusion
Data quality was negatively impacted by darker skin pigmentation for all tested devices, but skin pigmentation (ITA°) only increased photoplethysmographic HR measurement error to a small degree (~ 1 bpm) for the SlateSafety device.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Exercise Science, Mercer University, Macon, GA, USA. mulholland_a@mercer.edu.
- Department of Kinesiology, The University of Alabama, Tuscaloosa, AL, USA. mulholland_a@mercer.edu.
- Department of Kinesiology, The University of Alabama, Tuscaloosa, AL, USA.
Zitieren
APA
Mulholland, A. M., MacDonald, H. V., Aguiar, E. J., & Wingo, J. E. (2026). Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise. European Journal of Applied Physiology, 126(2), 1057-1066. https://doi.org/10.1007/s00421-025-05977-x
Vancouver
Mulholland AM, MacDonald HV, Aguiar EJ, Wingo JE. Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise. European Journal of Applied Physiology. 2026;126(2):1057-1066. doi:10.1007/s00421-025-05977-x
BibTeX
@article{Mulholland2026,
author = {Mulholland, Anne M and MacDonald, Hayley V and Aguiar, Elroy J and Wingo, Jonathan E},
title = {Influence of skin pigmentation on the accuracy and data quality of photoplethysmographic heart rate measurement during exercise},
journal = {European Journal of Applied Physiology},
year = {2026},
volume = {126},
number = {2},
pages = {1057--1066},
doi = {10.1007/s00421-025-05977-x}
}
EUVIMED