Fachartikel in der Zeitschrift European Radiology, 2024. Verfasst von Guillaume Fahrni und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, Cardiothoracic and Vascular Division, Lausanne University Hospital and University of Lausanne,…. Hauptthemen laut MEDLINE: Coronary Stenosis, Computed Tomography Angiography, Coronary Angiography, Tomography, X-Ray Computed. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EUROPE European Research Council, European Research Council.
Abstract (Originalsprache)
Objectives
To assess the image quality of ultra-high-resolution (UHR) virtual monoenergetic images (VMIs) at 40 keV compared to 70 keV, using spectral photon-counting CT (SPCCT) and dual-layer dual-energy CT (DECT) for coronary computed tomography angiography (CCTA). Methods and materials In this prospective IRB-approved study, 26 high-risk patients were included. CCTA was performed both with an SPCCT in UHR mode and with one of two DECT scanners (iQOn or CT7500) within 3 days. 40 keV and 70 keV VMIs were reconstructed for both modalities. Stenoses, blooming artefacts, and image quality were compared between all four reconstructions.
Results
Twenty-six patients (4 women [15%]) and 28 coronary stenoses (mean stenosis of 56% ± 16%) were included. 40 keV SPCCT gave an overall higher quality score (5 [5, 5]) than 70 keV SPCCT (5 [4, 5], 40 keV DECT (4 [3, 4]) and 70 keV SPCCT (4 [4, 5]), p < 0.001). Less variability in stenosis measurement was found with SPCCT between 40 keV and 70 keV (bias: −1% ± 3%, LoA: 6%) compared with DECT (−6% ± 8%, LoA 16%). 40 keV SPCCT vs 40 keV DECT showed a −3% ± 6% bias, whereas 40 keV SPCCT vs 70 keV DECT showed a −8% ± 6% bias. From 70 keV to 40 keV, blooming artefacts did not increase with SPCCT (mean +2% ± 5%, p = 0.136) whereas they increased with DECT (mean +7% ± 6%, p = 0.005).
Conclusion
UHR 40 keV SPCCT VMIs outperformed 40 keV and 70 keV DECT VMIs for assessing coronary artery stenoses, with no impairment compared to 70 keV SPCCT VMIs. Key Points Question Use of low virtual mono-energetic images at 40 keV using spectral dual-energy and photon-counting CT systems is not yet established for diagnosing coronary artery stenosis .
Findings
UHR 40 keV SPCCT enhances diagnostic accuracy in coronary artery assessment . Clinical relevance By combining spectral sensitivity with lower virtual mono-energetic imaging and ultra-high spatial resolution, SPCCT enhances coronary artery assessment, potentially leading to more accurate diagnoses and better patient outcomes in cardiovascular imaging .
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Diagnostic and Interventional Radiology, Cardiothoracic and Vascular Division, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.
- University of Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint Etienne, CNRS, Inserm, CREATIS, Villeurbanne, France.
- Department of Cardiovascular and Thoracic Radiology, Louis Pradel Hospital, Hospices Civils de Lyon, Bron, France.
- Philips Healthcare, Suresnes, France.
- Department of Radiology, CHU Nîmes, University Montpellier, Medical Imaging Group Nîmes, Nîmes, France.
- University of Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint Etienne, CNRS, Inserm, CREATIS, Villeurbanne, France. salim.si-mohamed@chu-lyon.fr.
Förderung
- HORIZON EUROPE European Research Council
- European Research Council
Zitieren
APA
Fahrni, G., Boccalini, S., Lacombe, H., de Oliveira, F., Houmeau, A., Francart, F., Villien, M., Rotzinger, D. C., Robert, A., Douek, P., & Si-Mohamed, S. A. (2024). Ultra-high-resolution 40 keV virtual monoenergetic imaging using spectral photon-counting CT in high-risk patients for coronary stenoses. European Radiology. https://doi.org/10.1007/s00330-024-11237-x
Vancouver
Fahrni G, Boccalini S, Lacombe H, de Oliveira F, Houmeau A, Francart F, et al. Ultra-high-resolution 40 keV virtual monoenergetic imaging using spectral photon-counting CT in high-risk patients for coronary stenoses. European Radiology. 2024. doi:10.1007/s00330-024-11237-x
BibTeX
@article{Fahrni2024,
author = {Fahrni, Guillaume and Boccalini, Sara and Lacombe, Hugo and de Oliveira, Fabien and Houmeau, Angèle and Francart, Florie and Villien, Marjorie and Rotzinger, David C and Robert, Antoine and Douek, Philippe and Si-Mohamed, Salim A},
title = {Ultra-high-resolution 40 keV virtual monoenergetic imaging using spectral photon-counting CT in high-risk patients for coronary stenoses},
journal = {European Radiology},
year = {2024},
doi = {10.1007/s00330-024-11237-x}
}
EUVIMED