Prospective comparison of ultra-high-resolution photon-counting detector CT coronary angiography versus conventional CT
British Journal of Radiology · 2026
DOI: 10.1093/bjr/tqaf220 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift British Journal of Radiology, 2026. Verfasst von Cynthia H McCollough und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States. Hauptthemen laut MEDLINE: Computed Tomography Angiography, Coronary Angiography, Coronary Stenosis. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, NIH HHS, Siemens Healthcare.
Abstract (Originalsprache)
Objectives
Determine differences in quantitative stenosis severity measurements for ultra-high-resolution (UHR) photon-counting detector (PCD) coronary CT angiography (cCTA) relative to energy-integrating detector (EID) cCTA in a large patient cohort.
Methods
Adult participants seen between November 2022 and March 2023 underwent a clinical dual-source EID-CT cCTA and a research dual-source PCD-CT cCTA on the same day. Percent diameter stenosis (PDS) was measured, and stenosis severity was assigned based on the PDS of the most severe lesion per patient to determine a coronary artery disease reporting and data system (CAD-RADS) score. Agreement between EID-CT and PCD-CT for PDS and CAD-RADS was determined.
Results
A total of 112 participants were enrolled, yielding 82 subjects with at least 1 stenosis at PCD-CT (mean age, 68.0 ± 10.8 years; 69.5% [57/82] males). A total of 177 paired stenoses were quantified. The percent decrease in mean PDS from PCD-CT (24.3%) to EID-CT (29.4%) was 17.3% (median difference in PDS: −5.0%, 95% CI: −5.7% to −4.0%). The assigned CAD-RADS score changed in 31/82 subjects. In 11/31 subjects, the most severe stenosis was missed at EID-CT due to partial volume averaging of a small calcification, yet PCD clearly identified stenosis at those locations. In 20/31 subjects, the CAD-RADS score decreased at PCD-CT due to the decreased calcium blooming resulting from the improved spatial resolution.
Conclusions
UHR PCD cCTA decreases quantitative measures of stenosis severity when a stenosis is identified at EID, leading to decreases in CAD-RADS assignments. Advances in knowledge With its improved spatial resolution, PCD-CT identifies stenoses missed on EID-CT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Mayo Clinic, Rochester, MN 55905, United States.
- Department of Health Science Research, Mayo Clinic, Jacksonville, FL 32224, United States.
ORCID-Profile: Cynthia H McCollough, Kishore Rajendran, Shuai Leng
Förderung
- NIBIB NIH HHS
- NIH HHS
- Siemens Healthcare
- National Institutes of Health
Zitieren
APA
McCollough, C. H., Davis, A. K., Weber, N. M., Kasten, H. J., Melka, E. F., Rajendran, K., Carter, R. E., & Leng, S. (2026). Prospective comparison of ultra-high-resolution photon-counting detector CT coronary angiography versus conventional CT. British Journal of Radiology. https://doi.org/10.1093/bjr/tqaf220
Vancouver
McCollough CH, Davis AK, Weber NM, Kasten HJ, Melka EF, Rajendran K, et al. Prospective comparison of ultra-high-resolution photon-counting detector CT coronary angiography versus conventional CT. British Journal of Radiology. 2026. doi:10.1093/bjr/tqaf220
BibTeX
@article{McCollough2026,
author = {McCollough, Cynthia H and Davis, Aeden K and Weber, Nikkole M and Kasten, Holly J and Melka, Elnata F and Rajendran, Kishore and Carter, Rickey E and Leng, Shuai},
title = {Prospective comparison of ultra-high-resolution photon-counting detector CT coronary angiography versus conventional CT},
journal = {British Journal of Radiology},
year = {2026},
doi = {10.1093/bjr/tqaf220}
}
EUVIMED