A preliminary validation study showing comparable accuracy between Computed Tomography based and imageless portable navigation systems when both systems are used
International Orthopaedics · 2026
DOI: 10.1007/s00264-026-06921-0 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift International Orthopaedics, 2026. Verfasst von Yuichi Miura und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Osaka Metropolitan University, Osaka, Japan. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Tomography, X-Ray Computed, Acetabulum, Surgical Navigation Systems. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben.
Abstract (Originalsprache)
Purpose
Accurate acetabular cup placement is essential in total hip arthroplasty (THA). We hypothesized that the newly introduced Computed Tomography (CT)-based portable navigation system would demonstrate accuracy comparable to that of the imageless portable navigation system. The aim of this study was to compare cup placement accuracy between the CT-based and imageless portable navigation systems of the same platform in THA performed in the lateral decubitus position.
Methods
This retrospective cross-sectional study included 36 patients who underwent primary THA via a direct lateral approach in the lateral decubitus position. In all cases, both imageless and CT-based portable navigation systems were used concurrently. Postoperative cup alignment was evaluated using three-dimensional CT (3D-CT). The primary outcome was the absolute error in cup inclination and anteversion, defined as the difference between intraoperative navigation values and postoperative 3D-CT measurements in the functional. Secondary outcomes included outlier rates and registration success rates.
Results
No statistically significant differences were observed between the imageless and CT-based portable navigation systems in the mean absolute error for inclination (2.2 ± 1.8° vs. 2.3 ± 1.8°, p = 0.93) or anteversion (2.3 ± 2.3° vs. 2.6 ± 2.5°, p = 0.41). There were no significant differences in outlier proportions. The registration success rate was 92% (36/39) due to three technical failures.
Conclusion
In this preliminary study, the CT-based portable navigation system demonstrated cup placement accuracy comparable to that of the imageless portable navigation system. Although the CT-based system may provide additional spatial information intraoperatively, its impact on clinical outcomes remains unclear and requires further longitudinal investigation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedic Surgery, Osaka Metropolitan University, Osaka, Japan.
- Department of Orthopedic Surgery, Osaka Metropolitan University, Osaka, Japan. ueyama.b@gmail.com.
ORCID-Profile: Hideki Ueyama
Zitieren
APA
Miura, Y., Ueyama, H., Sugama, R., Minoda, Y., Ohyama, Y., Masuda, S., Kirihigashi, R., & Terai, H. (2026). A preliminary validation study showing comparable accuracy between Computed Tomography based and imageless portable navigation systems when both systems are used. International Orthopaedics. https://doi.org/10.1007/s00264-026-06921-0
Vancouver
Miura Y, Ueyama H, Sugama R, Minoda Y, Ohyama Y, Masuda S, et al. A preliminary validation study showing comparable accuracy between Computed Tomography based and imageless portable navigation systems when both systems are used. International Orthopaedics. 2026. doi:10.1007/s00264-026-06921-0
BibTeX
@article{Miura2026,
author = {Miura, Yuichi and Ueyama, Hideki and Sugama, Ryo and Minoda, Yukihide and Ohyama, Yohei and Masuda, Sho and Kirihigashi, Ryosuke and Terai, Hidetomi},
title = {A preliminary validation study showing comparable accuracy between Computed Tomography based and imageless portable navigation systems when both systems are used},
journal = {International Orthopaedics},
year = {2026},
doi = {10.1007/s00264-026-06921-0}
}
EUVIMED