Evaluating the Accuracy of a Computed Tomography-Based Mixed-Reality Navigation Tool for Acetabular Component Positioning in Total Hip Arthroplasty

Eric S Dilbone, Alexander F Heimann, Justin Leal, Sean P Ryan, Samuel S Wellman

The Journal of Arthroplasty · 2025-08 · 40 (8) · S179-S185

DOI: 10.1016/j.arth.2025.02.003 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift The Journal of Arthroplasty, 2025-08. Verfasst von Eric S Dilbone und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Duke University, Durham, North Carolina. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Acetabulum, Tomography, X-Ray Computed, Surgery, Computer-Assisted. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 48 Literaturangaben. Gefördert von: Duke University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Robotics and navigation systems have improved the accuracy and precision of acetabular component placement in total hip arthroplasty (THA). The purpose of this study was to determine if the intraoperative use of a novel mixed-reality (MR) navigation system resulted in accurate acetabular component placement with minimal outliers. METHODS: A series of 79 patients who underwent THA performed by a single surgeon using a novel MR navigation system were retrospectively reviewed. The preoperative planned acetabular component operative inclination (OI) and tilt-adjusted operative anteversion (OA) were collected for each case. At the six-week follow-up, each patient underwent three-dimensional EOS imaging, which was used to measure the acetabular component placement via validated measuring software. Patients who had bilateral THAs or three-dimensional images that did not have critical bony landmarks visible for measurement were excluded from the study. Absolute error was calculated for both inclination and anteversion. Outliers were defined as any cup position that was outside a range of ± 10 degrees from the preoperatively planned target. A total of 39 patients met the inclusion criteria. RESULTS: The mean planned OI for each patient was 40.8 (range, 40 to 41) degrees, and the mean measured postoperative OI was 40.4 (range, 37 to 46) degrees (P = 0.40), resulting in a mean absolute error of 1.8 (range, zero to five) degrees. The mean planned OA was 30.2 (range, 25 to 35) degrees, and the mean measured postoperative OA was 31.1 (range, 27 to 42) degrees (P = 0.11), resulting in a mean absolute error of 2.0 (range, zero to six) degrees. There were no outliers in either OI or OA. CONCLUSIONS: The results of this study suggest that the use of MR navigation during THA results in accurate postoperative acetabular component inclination and anteversion relative to the preoperative plan without outliers.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipAcetabulumTomography, X-Ray ComputedSurgery, Computer-AssistedSurgical Navigation SystemsHumansFemaleMaleRetrospective StudiesMiddle AgedAgedHip ProsthesisAdultImaging, Three-DimensionalAged, 80 and over

Stichwörter der Autor:innen

anteversionaugmented-reality (AR)inclinationmixed-reality (MR)total hip arthroplasty (THA)

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Duke University, Durham, North Carolina.
  • Department of Orthopaedic Surgery and Traumatology, HFR Fribourg-Cantonal Hospital, University of Fribourg, Fribourg, Switzerland.

Förderung

  • Duke University

Zitieren

APA

Dilbone, E. S., Heimann, A. F., Leal, J., Ryan, S. P., & Wellman, S. S. (2025). Evaluating the Accuracy of a Computed Tomography-Based Mixed-Reality Navigation Tool for Acetabular Component Positioning in Total Hip Arthroplasty. The Journal of Arthroplasty, 40(8), S179-S185. https://doi.org/10.1016/j.arth.2025.02.003

Vancouver

Dilbone ES, Heimann AF, Leal J, Ryan SP, Wellman SS. Evaluating the Accuracy of a Computed Tomography-Based Mixed-Reality Navigation Tool for Acetabular Component Positioning in Total Hip Arthroplasty. The Journal of Arthroplasty. 2025;40(8):S179-S185. doi:10.1016/j.arth.2025.02.003

BibTeX
@article{Dilbone2025,
  author = {Dilbone, Eric S and Heimann, Alexander F and Leal, Justin and Ryan, Sean P and Wellman, Samuel S},
  title = {Evaluating the Accuracy of a Computed Tomography-Based Mixed-Reality Navigation Tool for Acetabular Component Positioning in Total Hip Arthroplasty},
  journal = {The Journal of Arthroplasty},
  year = {2025},
  volume = {40},
  number = {8},
  pages = {S179--S185},
  doi = {10.1016/j.arth.2025.02.003}
}
Kostenlos starten →