EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Accuracy of two different imageless navigation systems for leg length and global offset change in total hip arthroplasty: A comparison using two-dimensional radiographic and three-dimensional CT-based evaluation

Shine Tone, Yohei Naito, Gai Kobayashi, Hironori Unno, Akihiro Sudo, Masahiro Hasegawa

International Orthopaedics · 2026

DOI: 10.1007/s00264-026-06849-5 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift International Orthopaedics, 2026. Verfasst von Shine Tone und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Mie University, Graduate School of Medicine, Tsu, Japan. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Tomography, X-Ray Computed, Leg Length Inequality, Surgery, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Mie University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aimed to evaluate the accuracy of two imageless navigation systems for restoring leg length change (LLC) and global offset change (GOC) in total hip arthroplasty (THA) using two-dimensional (2D) radiographic and three-dimensional computed tomography (3D CT)-based assessment methods.

Methods

Patients undergoing primary cementless THA were divided into two groups based on the imageless navigation system used: a large-console group ( n = 120) and a portable handheld group ( n = 83). Intraoperative navigation measurements of the LLC and GOC were compared with values derived from preoperative and postoperative assessments, and absolute measurement errors were calculated. Accuracy was evaluated using 2D radiographic and 3D CT-based measurements. Between-system differences and discrepancies between 2 and 3D assessment methods were analyzed.

Results

Absolute LLC error in the large-console group was 2.7 ± 3.3 mm on 2D radiographic evaluation and 2.5 ± 3.3 mm on 3D CT-based evaluation, compared with 2.9 ± 2.7 mm and 3.0 ± 2.8 mm, respectively, in the portable handheld group. LLC error was significantly lower in the large-console group on 3D evaluation (p = 0.004). Absolute GOC error did not differ significantly between groups. No differences were observed between 2 and 3D evaluations for LLC, whereas most GOC-related parameters differed significantly between methods.

Conclusion

Imageless navigation systems achieved favorable accuracy for LLC and GOC in THA. While radiographic assessment is sufficient for evaluating leg length, 3D CT-based evaluation provides a more consistent and less position-dependent assessment of global offset.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipTomography, X-Ray ComputedLeg Length InequalitySurgery, Computer-AssistedSurgical Navigation SystemsHumansFemaleImaging, Three-DimensionalMaleMiddle AgedAgedHip JointAdult

Stichwörter der Autor:innen

Global offset changeImageless navigationLarge-consoleLeg length changePortableTotal hip arthroplasty

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Mie University, Graduate School of Medicine, Tsu, Japan.
  • Department of Orthopaedic Surgery, Mie University, Graduate School of Medicine, Tsu, Japan. yo-yo@med.mie-u.ac.jp.
  • Yonaha Okanoue Hospital, Kuwana, Japan.

Förderung

  • Mie University

Zitieren

APA

Tone, S., Naito, Y., Kobayashi, G., Unno, H., Sudo, A., & Hasegawa, M. (2026). Accuracy of two different imageless navigation systems for leg length and global offset change in total hip arthroplasty: A comparison using two-dimensional radiographic and three-dimensional CT-based evaluation. International Orthopaedics. https://doi.org/10.1007/s00264-026-06849-5

Vancouver

Tone S, Naito Y, Kobayashi G, Unno H, Sudo A, Hasegawa M. Accuracy of two different imageless navigation systems for leg length and global offset change in total hip arthroplasty: A comparison using two-dimensional radiographic and three-dimensional CT-based evaluation. International Orthopaedics. 2026. doi:10.1007/s00264-026-06849-5

BibTeX
@article{Tone2026,
  author = {Tone, Shine and Naito, Yohei and Kobayashi, Gai and Unno, Hironori and Sudo, Akihiro and Hasegawa, Masahiro},
  title = {Accuracy of two different imageless navigation systems for leg length and global offset change in total hip arthroplasty: A comparison using two-dimensional radiographic and three-dimensional CT-based evaluation},
  journal = {International Orthopaedics},
  year = {2026},
  doi = {10.1007/s00264-026-06849-5}
}
Kostenlos starten →