Accuracy of a novel, automated 2D-3D registration software for measuring cup position in total hip arthroplasty
International Journal of Computer Assisted Radiology and Surgery · 11.05.2025 · 20 (10) · 2043-2051
DOI: 10.1007/s11548-025-03389-y ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, online erschienen am 11.05.2025. Verfasst von Jevan Arulampalam und 7 weiteren Autor:innen. Beteiligte Einrichtung: Corin Group, Raynham, MA, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Imaging, Three-Dimensional, Software, Tomography, X-Ray Computed. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 40 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: This study evaluated the accuracy of an automated 2D-3D registration software for matching preoperative 3D models of the pelvis and acetabular component to intraoperative 2D fluoroscopy images in total hip arthroplasty (THA). METHODS: We developed a 2D-3D registration software that registers a 3D model of the pelvis from preoperative CT and a 3D model of the acetabular implant to intraoperative fluoroscopic imaging, thereby calculating the implant position relative to the 3D pelvic reference frame. A total of 145 datasets were used including 65 digitally reconstructed radiographs, 20 dry bone phantoms datasets and 60 clinical datasets with preoperative CT and intraoperative fluoroscopy imaging. Achieved acetabular positions from the clinical images were determined from post-operative CT using a 3D/3D registration method. Accuracy was assessed by comparing the calculated acetabular position from the 2D-3D software to the ground truth data. RESULTS: Mean absolute difference between ground truth and the 2D-3D software was 1.9° [signed error range: -4.4, 4.8] for inclination, 1.5° [-7.3, 4.1] for anteversion, 1.6 mm [-5, 3.8] for cup height and 1.8 mm [-7.3, 4.1] for depth across all datasets. In total, 100% of inclination results and 98% of anteversion results were within 5° while 90% of height and 81% of depth results were within 3 mm. CONCLUSION: We validated the accuracy of an automated 2D-3D registration software for use in THA. While our method requires preoperative data from CT, the results are comparable to robotics and image-based navigation, and present a promising, simple technology that can be easily integrated into an operating room for THA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Corin Group, Raynham, MA, USA.
- Corin Group, Raynham, MA, USA. Christopher.Plaskos@coringroup.com.
- School of Aerospace, Mechanical and Mechatronic Engineering, University of Sydney, Sydney, Australia.
- SAN Hospital, Wahroonga, NSW, Australia.
- Norwest Private Hospital, Bella Vista, NSW, Australia.
- Department of Orthopaedics, Prince of Wales, St Vincent's Private and Mater Hospitals, Sydney, NSW, Australia.
Zitieren
APA
Arulampalam, J., Ehlke, M. F., Plaskos, C., Li, Q., Stambouzou, C. Z., Sullivan, J. A., Solomon, M. I., & Slotkin, E. M. (2025). Accuracy of a novel, automated 2D-3D registration software for measuring cup position in total hip arthroplasty. International Journal of Computer Assisted Radiology and Surgery, 20(10), 2043-2051. https://doi.org/10.1007/s11548-025-03389-y
Vancouver
Arulampalam J, Ehlke MF, Plaskos C, Li Q, Stambouzou CZ, Sullivan JA, et al. Accuracy of a novel, automated 2D-3D registration software for measuring cup position in total hip arthroplasty. International Journal of Computer Assisted Radiology and Surgery. 2025;20(10):2043-2051. doi:10.1007/s11548-025-03389-y
BibTeX
@article{Arulampalam2025,
author = {Arulampalam, Jevan and Ehlke, Moritz F and Plaskos, Christopher and Li, Qing and Stambouzou, Catherine Z and Sullivan, James A and Solomon, Michael I and Slotkin, Eric M},
title = {Accuracy of a novel, automated 2D-3D registration software for measuring cup position in total hip arthroplasty},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2025},
volume = {20},
number = {10},
pages = {2043--2051},
doi = {10.1007/s11548-025-03389-y}
}
EUVIMED