Stress shielding effect after total hip arthroplasty varies between combinations of stem design and stiffness—a comparing biomechanical finite element analysis
International Orthopaedics · 2023-08 · 47 (8) · 1981-1987
DOI: 10.1007/s00264-023-05825-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Orthopaedics, online erschienen am 03.06.2023. Verfasst von Rene Burchard und 3 weiteren Autor:innen. Beteiligte Einrichtung: University of Marburg, Marburg, Germany. burcharr@staff.uni-marburg.de. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis. Die Arbeit wurde bisher 43-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Philipps-Universität Marburg.
Abstract (Originalsprache)
Purpose
Total hip arthroplasty (THA) has become a highly frequent orthopaedic procedure. Multiple approaches have been made to design the femoral component for THA with a mechanical behaviour as close as possible to a natural femur. The aim of this study was to compare different combinations of design and biomechanical properties of THA prostheses and their impact on stress shielding of the periprosthetic bone.
Methods
Virtual implantation of different stem designs (straight standard stem, straight short stem, anatomical short stem) by finite element analysis based on in vivo data from computer tomography was performed. For each stem, three grades of stiffness were generated, followed by a strain analysis.
Results
Reduction of stem stiffness led to less stress shielding. Implantation of an anatomical short-stem prosthesis with low stiffness provided the most physiological strain-loading effect ( p < 0.001).
Conclusion
A combination of a short and an anatomically designed stem with a low stiffness might provide a more physiological strain transfer during THA. Biomechanical properties of the femoral component for THA should be considered as a multifactorial function of dimensions, design, and stiffness.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Marburg, Marburg, Germany. burcharr@staff.uni-marburg.de.
- Department of Orthopedics and Trauma Surgery, University of Giessen and Marburg, Marburg, Germany. burcharr@staff.uni-marburg.de.
- Department of Orthopedics and Trauma Surgery, Lahn-Dill-Kliniken, Rotebergstr. 2, 35683, Dillenburg, Germany. burcharr@staff.uni-marburg.de.
- Department of Anesthesiology and Intensive Care Medicine, Ulm University Hospital, Ulm, Germany.
- FOM University of Applied Sciences, Essen, Germany.
- University of Marburg, Marburg, Germany.
ORCID-Profile: Rene Burchard
Förderung
- Philipps-Universität Marburg
Zitieren
APA
Burchard, R., Graw, J. A., Soost, C., & Schmitt, J. (2023). Stress shielding effect after total hip arthroplasty varies between combinations of stem design and stiffness—a comparing biomechanical finite element analysis. International Orthopaedics, 47(8), 1981-1987. https://doi.org/10.1007/s00264-023-05825-7
Vancouver
Burchard R, Graw JA, Soost C, Schmitt J. Stress shielding effect after total hip arthroplasty varies between combinations of stem design and stiffness—a comparing biomechanical finite element analysis. International Orthopaedics. 2023;47(8):1981-1987. doi:10.1007/s00264-023-05825-7
BibTeX
@article{Burchard2023,
author = {Burchard, Rene and Graw, Jan A and Soost, Christian and Schmitt, Jan},
title = {Stress shielding effect after total hip arthroplasty varies between combinations of stem design and stiffness—a comparing biomechanical finite element analysis},
journal = {International Orthopaedics},
year = {2023},
volume = {47},
number = {8},
pages = {1981--1987},
doi = {10.1007/s00264-023-05825-7}
}
EUVIMED