A novel design of hip-stem with reduced strain-shielding
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · 2024
DOI: 10.1177/09544119241244537 ↗Kurz gefasst
Fachartikel in der Zeitschrift Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2024. Verfasst von Tanmoy Loha und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Indian Institute of Engineering Science and Technology, Shibpur, Howrah, India. Hauptthemen laut MEDLINE: Prosthesis Design, Hip Prosthesis, Stress, Mechanical, Finite Element Analysis. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 66 Literaturangaben. Gefördert von: Wellcome Trust and the EPSRC.
Abstract (Originalsprache)
The use of uncemented stems in hip arthroplasty has been increasing, even in osteoporotic patients. The major concerns of uncemented hip-stems, however, are peri-prosthetic fracture, thigh pain, and proximal femoral stress-/strain-shielding. In this study, a novel design of uncemented hip-stem is proposed that will reduce such concerns, improve osseointegration, and benefit both osteoporotic and arthritic patients. The stem has a central titanium alloy core surrounded by a set of radial buttresses that are partly porous titanium, as is the stem tip. The aim of the study was to investigate the mechanical behaviour of the proposed partly-porous design, examining load transfer in the short-term, and comparing its strain-shielding behaviour with a solid metal implant. The long-term effect of implant-induced bone remodelling was also simulated. Computed tomography based three-dimensional finite element models of an intact proximal femur, and the same femur implanted with the proposed design, were developed. Peak hip contact and major muscle forces corresponding to level-walking and stair climbing were applied. The proposed partly-porous design had approximately 50% lower strain-shielding than the solid-metal counterpart. Results of bone remodelling simulation indicated that only 16% of the total bone volume is subjected to reduction of bone density. Strain concentrations were observed in the bone around the stem-tip for both solid and porous implants; however, it was less prominent for the porous design. Lower strain-shielding and reduced bone resorption are advantageous for long-term fixation, and the reduced strain concentration around the stem-tip indicates a lower risk of peri-prosthetic fracture.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Indian Institute of Engineering Science and Technology, Shibpur, Howrah, India.
- Department of Mechanical Engineering, Imperial College London, London, UK.
ORCID-Profile: Tanmoy Loha, Bidyut Pal
Förderung
- Wellcome Trust and the EPSRC
Zitieren
APA
Loha, T., Bhattacharya, R., Pal, B., & Amis, A. A. (2024). A novel design of hip-stem with reduced strain-shielding. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. https://doi.org/10.1177/09544119241244537
Vancouver
Loha T, Bhattacharya R, Pal B, Amis AA. A novel design of hip-stem with reduced strain-shielding. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2024. doi:10.1177/09544119241244537
BibTeX
@article{Loha2024,
author = {Loha, Tanmoy and Bhattacharya, Rounak and Pal, Bidyut and Amis, Andrew A},
title = {A novel design of hip-stem with reduced strain-shielding},
journal = {Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine},
year = {2024},
doi = {10.1177/09544119241244537}
}
EUVIMED