EUVIMEDEUVIMED entdecken
Open Access

Efficient and complete fat suppression in two-point Dixon MR imaging: sequence considerations and assessment in lumbar plexus imaging

Sheng-Qing Lin, Durga Udayakumar, Sebastian Fonseca, Neela Batthula, Flavio Duarte Silva, Avneesh Chhabra, Ananth J Madhuranthakam

Skeletal Radiology · 2026-09 · 55 (9) · 2351-2361

DOI: 10.1007/s00256-026-05174-7 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Skeletal Radiology, online erschienen am 03.03.2026. Verfasst von Sheng-Qing Lin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, UT Southwestern Medical Center, Dallas, TX, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Adipose Tissue, Lumbosacral Plexus, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, Cancer Prevention and Research Institute of Texas, National Cancer Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: Two-point Dixon MRI is widely used for lumbar plexus imaging but often yields incomplete fat suppression due to neglecting fat's multi-peak spectrum. This study developed and evaluated a post-processing algorithm that accounts for the multi-peak fat spectrum to achieve more effective fat suppression. METHODS: A novel Dark-fat processing algorithm was developed to enhance fat suppression in water-only MR images from standard two-point Dixon reconstruction. It was evaluated using a custom fat/water phantom, prospectively in a healthy volunteer, and retrospectively in 30 patients who underwent clinical lumbar plexus imaging between March and October 2023. Imaging was done using FFE (or GRE)-Dixon, TSE (or FSE)-Dixon, and mDixon-Quant. Quantitative metrics (SNR, apparent SNR (aSNR), and apparent CNR (aCNR)) and qualitative assessments by two radiologists were analyzed. Statistical significance was determined using Welch's t-test (p 1) and muscle (p < 0.01, effect size = 0.81); and (iv) did not alter aCNR between nerve and muscle tissue (p = 0.27, effect size = 0.29). CONCLUSION: Dark-fat processing enables efficient and complete fat suppression in two-point Dixon acquisitions, especially TSE-Dixon, significantly enhancing lumbar plexus image quality. When implemented offline and as a retrospective method, it requires no change to scan time and can be applied to any two-point Dixon dataset.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingAdipose TissueLumbosacral PlexusImage Interpretation, Computer-AssistedHumansFemaleAlgorithmsMaleAdultRetrospective StudiesPhantoms, ImagingMiddle Aged

Stichwörter der Autor:innen

Dark-fat processingDixon imagingFFE-DixonFat suppressionLumbar plexusTSE-Dixon

Autor:innen und Institutionen

  • Department of Radiology, UT Southwestern Medical Center, Dallas, TX, USA.
  • Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, USA.
  • Present address: Department of Radiology, Mayo Clinic, 200 First Street SW, Rochester, MN, 55905, USA.
  • Department of Orthopedic Surgery, UT Southwestern Medical Center, Dallas, TX, USA.
  • Department of Radiology, UT Southwestern Medical Center, Dallas, TX, USA. ananth@mayo.edu.
  • Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, USA. ananth@mayo.edu.

Förderung

  • NCI NIH HHS
  • Cancer Prevention and Research Institute of Texas
  • National Cancer Institute

Zitieren

APA

Lin, S. Q., Udayakumar, D., Fonseca, S., Batthula, N., Silva, F. D., Chhabra, A., & Madhuranthakam, A. J. (2026). Efficient and complete fat suppression in two-point Dixon MR imaging: sequence considerations and assessment in lumbar plexus imaging. Skeletal Radiology, 55(9), 2351-2361. https://doi.org/10.1007/s00256-026-05174-7

Vancouver

Lin SQ, Udayakumar D, Fonseca S, Batthula N, Silva FD, Chhabra A, et al. Efficient and complete fat suppression in two-point Dixon MR imaging: sequence considerations and assessment in lumbar plexus imaging. Skeletal Radiology. 2026;55(9):2351-2361. doi:10.1007/s00256-026-05174-7

BibTeX
@article{Lin2026,
  author = {Lin, Sheng-Qing and Udayakumar, Durga and Fonseca, Sebastian and Batthula, Neela and Silva, Flavio Duarte and Chhabra, Avneesh and Madhuranthakam, Ananth J},
  title = {Efficient and complete fat suppression in two-point Dixon MR imaging: sequence considerations and assessment in lumbar plexus imaging},
  journal = {Skeletal Radiology},
  year = {2026},
  volume = {55},
  number = {9},
  pages = {2351--2361},
  doi = {10.1007/s00256-026-05174-7}
}
Kostenlos starten →