Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 23.06.2024. Verfasst von Yavuz Muslu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Liver, Magnetic Resonance Imaging, Phantoms, Imaging, Algorithms. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 102 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Bayer HealthCare, NIH HHS, NIDDK NIH HHS.
Abstract (Originalsprache)
Purpose
Quantitative T 1 mapping has the potential to replace biopsy for noninvasive diagnosis and quantitative staging of chronic liver disease. Conventional T 1 mapping methods are confounded by fat and inhomogeneities, resulting in unreliable T 1 estimations. Furthermore, these methods trade off spatial resolution and volumetric coverage for shorter acquisitions with only a few images obtained within a breath‐hold. This work proposes a novel, volumetric (3D), free‐breathing T 1 mapping method to account for multiple confounding factors in a single acquisition. Theory and Methods Free‐breathing, confounder‐corrected T 1 mapping was achieved through the combination of non‐Cartesian imaging, magnetization preparation, chemical shift encoding, and a variable flip angle acquisition. A subspace‐constrained, locally low‐rank image reconstruction algorithm was employed for image reconstruction. The accuracy of the proposed method was evaluated through numerical simulations and phantom experiments with a T 1 /proton density fat fraction phantom at 3.0 T. Further, the feasibility of the proposed method was investigated through contrast‐enhanced imaging in healthy volunteers, also at 3.0 T.
Results
The method showed excellent agreement with reference measurements in phantoms across a wide range of T 1 values (200 to 1000 ms, slope = 0.998 (95% confidence interval (CI) [0.963 to 1.035]), intercept = 27.1 ms (95% CI [0.4 54.6]), r 2 = 0.996), and a high level of repeatability. In vivo imaging studies demonstrated moderate agreement (slope = 1.099 (95% CI [1.067 to 1.132]), intercept = −96.3 ms (95% CI [−82.1 to −110.5]), r 2 = 0.981) compared to saturation recovery‐based T 1 maps.
Conclusion
The proposed method produces whole‐liver, confounder‐corrected T 1 maps through simultaneous estimation of T 1 , proton density fat fraction, and in a single, free‐breathing acquisition and has excellent agreement with reference measurements in phantoms.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- GE HealthCare, Waukesha, Wisconsin, USA.
- Waisman Center, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Electrical and Computer Engineering, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA.
ORCID-Profile: Yavuz Muslu, Daiki Tamada, Nathan T Roberts, Ty A Cashen, Sagar Mandava, Steven R Kecskemeti, Diego Hernando, Scott B Reeder
Förderung
- Bayer HealthCare
- NIH HHS
- NIDDK NIH HHS
- NIBIB NIH HHS
- National Institutes of Health
- Bayer Diabetes Care
Zitieren
APA
Muslu, Y., Tamada, D., Roberts, N. T., Cashen, T. A., Mandava, S., Kecskemeti, S. R., Hernando, D., & Reeder, S. B. (2024). Free‐breathing, fat‐corrected T 1 mapping of the liver with stack‐of‐stars MRI , and joint estimation of T 1 , PDFF , R2*, and B1+. Magnetic Resonance in Medicine, 92(5), 1913-1932. https://doi.org/10.1002/mrm.30182
Vancouver
Muslu Y, Tamada D, Roberts NT, Cashen TA, Mandava S, Kecskemeti SR, et al. Free‐breathing, fat‐corrected T 1 mapping of the liver with stack‐of‐stars MRI , and joint estimation of T 1 , PDFF , R2*, and B1+. Magnetic Resonance in Medicine. 2024;92(5):1913-1932. doi:10.1002/mrm.30182
BibTeX
@article{Muslu2024,
author = {Muslu, Yavuz and Tamada, Daiki and Roberts, Nathan T and Cashen, Ty A and Mandava, Sagar and Kecskemeti, Steven R and Hernando, Diego and Reeder, Scott B},
title = {Free‐breathing, fat‐corrected T 1 mapping of the liver with stack‐of‐stars MRI , and joint estimation of T 1 , PDFF , R2*, and B1+},
journal = {Magnetic Resonance in Medicine},
year = {2024},
volume = {92},
number = {5},
pages = {1913--1932},
doi = {10.1002/mrm.30182}
}
EUVIMED