EUVIMEDEUVIMED entdecken
Crossref

Ion chromatography for inorganic phosphate in human serum: a candidate reference measurement procedure and commutable certified reference material

Sin Yee Ng, Ho Wah Leung, Richard Y C Shin, Qinde Liu, Tang Lin Teo, Sunil Kumar Sethi

Analytical and Bioanalytical Chemistry · 07.11.2026

DOI: 10.1007/s00216-026-06895-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical and Bioanalytical Chemistry, online erschienen am 07.11.2026. Verfasst von Sin Yee Ng und 5 weiteren Autor:innen. Beteiligte Einrichtung: Chemical Metrology Laboratory, Chemical Metrology Division, Applied Sciences Group, Health Sciences Authority of Singapore, 1 Science Park Road, #01-…. Die Arbeit nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Accurate measurement of inorganic phosphate in human serum is critical for clinical diagnostics, yet the metrological infrastructure supporting this analyte remains limited, with only one reference measurement procedure (RMP) listed by the Joint Committee for Traceability in Laboratory Medicine (JCTLM) and no certified reference material (CRM) available for inorganic phosphate in human serum. This study addresses both gaps through the development and validation of an ion chromatography (IC) method with suppressed conductivity detection for the quantification of inorganic phosphate in human serum, with calibration standards traceable to SI units via NIST SRM 3186. Excellent linearity (R2 ≥ 0.999), precision (coefficient of variation, CV < 1%), and spike recoveries close to 100% were achieved, with expanded measurement uncertainties below 2% (k = 2), fulfilling optimal-level analytical performance specifications derived from biological variation data over a clinically relevant inorganic phosphate range in serum, supporting the suitability of this IC method as a candidate RMP. The method was applied to certify the inorganic phosphate concentration in frozen human serum (HRM-2023A) as 1.120 ± 0.023 mmol/L (k = 2), representing the first metrologically traceable CRM for inorganic phosphate in human serum. Commutability of HRM-2023A was confirmed against the Roche COBAS C311 and Abbott Alinity C clinical platforms using a regression-based approach with 95% prediction intervals. Good agreement with both routine analysers was demonstrated using 28 anonymised patient samples, with differences predominantly within the Royal College of Pathologists of Australasia (RCPA) acceptable limits of ± 8%. This work presents IC as a strong candidate RMP for inorganic phosphate in serum, supported by a commutable, SI-traceable CRM, addressing critical gaps in metrological infrastructure and supporting improved harmonisation of inorganic phosphate measurements in clinical laboratories.

Schlagwörter

Stichwörter der Autor:innen

Certified reference materialCommutabilityInorganic phosphateIon chromatographyMetrological traceabilityReference measurement procedure

Autor:innen und Institutionen

  • Chemical Metrology Laboratory, Chemical Metrology Division, Applied Sciences Group, Health Sciences Authority of Singapore, 1 Science Park Road, #01-05/06, The Capricorn, Singapore Science Park II, Singapore, 117528, Si…
  • Department of Laboratory Medicine, National University Hospital, Singapore, Singapore.

Zitieren

APA

Ng, S. Y., Leung, H. W., Shin, R. Y. C., Liu, Q., Teo, T. L., & Sethi, S. K. (2026). Ion chromatography for inorganic phosphate in human serum: a candidate reference measurement procedure and commutable certified reference material. Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-026-06895-3

Vancouver

Ng SY, Leung HW, Shin RYC, Liu Q, Teo TL, Sethi SK. Ion chromatography for inorganic phosphate in human serum: a candidate reference measurement procedure and commutable certified reference material. Analytical and Bioanalytical Chemistry. 2026. doi:10.1007/s00216-026-06895-3

BibTeX
@article{Ng2026,
  author = {Ng, Sin Yee and Leung, Ho Wah and Shin, Richard Y C and Liu, Qinde and Teo, Tang Lin and Sethi, Sunil Kumar},
  title = {Ion chromatography for inorganic phosphate in human serum: a candidate reference measurement procedure and commutable certified reference material},
  journal = {Analytical and Bioanalytical Chemistry},
  year = {2026},
  doi = {10.1007/s00216-026-06895-3}
}
Kostenlos starten →