EUVIMEDEUVIMED entdecken
Crossref

A liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of vancomycin in human serum and plasma

Sandra Fleischer, Neeraj Singh, Friederike Bauland, Andrea Geistanger, Judith Taibon, Elie Fux

Clinical Chemistry and Laboratory Medicine (CCLM) · 12.08.2026

DOI: 10.1515/cclm-2026-0509 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), online erschienen am 12.08.2026. Verfasst von Sandra Fleischer und 5 weiteren Autor:innen. Beteiligte Einrichtung: Roche Diagnostics GmbH, Penzberg, Germany. Die Arbeit nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Vancomycin is a glycopeptide antibiotic used to treat severe Gram-positive infections; its narrow therapeutic window makes therapeutic drug monitoring (TDM) critical. We report the development of a liquid chromatography-tandem mass spectrometry (LC-MS/MS) candidate reference measurement procedure (RMP) to quantify vancomycin in human plasma and serum.

Methods

Quantitative nuclear magnetic resonance (qNMR) spectroscopy was used to determine the absolute content of the reference material, ensuring direct traceability to the International System of Units (SI). Sample preparation used a protein precipitation protocol followed by high dilution. LC-MS/MS parameters, including selectivity, precision, accuracy, equivalence to the listed RMP, and measurement uncertainty (MU) were evaluated. MU determination was performed according to the guide to the expression of uncertainty in measurement (GUM).

Results

The RMP was validated over a measuring interval of 0.690–58.7 μmol/L (1.00–85.0 μg/mL). Intermediate precision and repeatability were

Schlagwörter

Stichwörter der Autor:innen

SI unitsliquid chromatography-tandem mass spectrometryqNMR characterizationreference measurement proceduretraceabilityvancomycin

Autor:innen und Institutionen

  • Roche Diagnostics GmbH, Penzberg, Germany.
  • Evidenze Germany GmbH, Essen, Germany.

Zitieren

APA

Fleischer, S., Singh, N., Bauland, F., Geistanger, A., Taibon, J., & Fux, E. (2026). A liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of vancomycin in human serum and plasma. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2026-0509

Vancouver

Fleischer S, Singh N, Bauland F, Geistanger A, Taibon J, Fux E. A liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of vancomycin in human serum and plasma. Clinical Chemistry and Laboratory Medicine (CCLM). 2026. doi:10.1515/cclm-2026-0509

BibTeX
@article{Fleischer2026,
  author = {Fleischer, Sandra and Singh, Neeraj and Bauland, Friederike and Geistanger, Andrea and Taibon, Judith and Fux, Elie},
  title = {A liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of vancomycin in human serum and plasma},
  journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
  year = {2026},
  doi = {10.1515/cclm-2026-0509}
}
Kostenlos starten →