Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry
Journal of Clinical Monitoring · 1991-10 · 7 (4) · 309-312
DOI: 10.1007/bf01619351 ↗At a glance
Fallbericht in Journal of Clinical Monitoring, 1991-10. Written by H M Sami and 2 co-authors. Institution: Department of Anesthesiology, Loyola University of Chicago, Maywood, IL 60153. Major MEDLINE topics: Central Venous Pressure, Intra-Aortic Balloon Pumping, Oximetry, Pulse. Cited 33 times so far (Crossref); 5 references.
Abstract (original language)
Venous pulsations are thought, but to our knowledge have never been demonstrated, to result in underestimation of oxygen saturation measured by pulse oximetry (SpO2); however, to our knowledge, this has never been demonstrated. A patient with ischemic cardiomyopathy requiring intra-aortic balloon counterpulsation (IABC) had to withstand a temporary cessation of IABC (period I), during which she had marked central venous pulsations and a pulse oximeter bias of -18%. After IABC was reinstituted (period II), the venous pulsations disappeared and the bias was reduced to 0%. Cardiac output, systemic vascular resistance, and temperature during both periods were comparable. Pulmonary vascular resistance, however, was much lower during period II. We postulate that the bias was caused by venous pulsations significantly contributing to the AC, or variable component, of light absorption.
Subjects
MeSH terms (MEDLINE)
Substances
Authors and institutions
- Department of Anesthesiology, Loyola University of Chicago, Maywood, IL 60153.
Cite
APA
Sami, H. M., Kleinman, B. S., & Lonchyna, V. A. (1991). Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry. Journal of Clinical Monitoring, 7(4), 309-312. https://doi.org/10.1007/bf01619351
Vancouver
Sami HM, Kleinman BS, Lonchyna VA. Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry. Journal of Clinical Monitoring. 1991;7(4):309-312. doi:10.1007/bf01619351
BibTeX
@article{Sami1991,
author = {Sami, H M and Kleinman, B S and Lonchyna, V A},
title = {Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry},
journal = {Journal of Clinical Monitoring},