EUVIMEDEUVIMED entdecken
Fallbericht

Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry

H M Sami, B S Kleinman, V A Lonchyna

Journal of Clinical Monitoring · 1991-10 · 7 (4) · 309-312

DOI: 10.1007/bf01619351 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Journal of Clinical Monitoring, 1991-10. Verfasst von H M Sami und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Loyola University of Chicago, Maywood, IL 60153. Hauptthemen laut MEDLINE: Central Venous Pressure, Intra-Aortic Balloon Pumping, Oximetry, Pulse. Die Arbeit wurde bisher 33-mal zitiert (Crossref) und nennt 5 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Venous pulsations are thought, but to our knowledge have never been demonstrated, to result in underestimation of oxygen saturation measured by pulse oximetry (SpO2); however, to our knowledge, this has never been demonstrated. A patient with ischemic cardiomyopathy requiring intra-aortic balloon counterpulsation (IABC) had to withstand a temporary cessation of IABC (period I), during which she had marked central venous pulsations and a pulse oximeter bias of -18%. After IABC was reinstituted (period II), the venous pulsations disappeared and the bias was reduced to 0%. Cardiac output, systemic vascular resistance, and temperature during both periods were comparable. Pulmonary vascular resistance, however, was much lower during period II. We postulate that the bias was caused by venous pulsations significantly contributing to the AC, or variable component, of light absorption.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Central Venous PressureIntra-Aortic Balloon PumpingOximetryPulseAgedCarbon DioxideCatheterization, PeripheralCatheterization, Swan-GanzCoronary DiseaseCounterpulsationFemoral ArteryHeart RateHumansMaleMonitoring, IntraoperativeOxygenPulmonary ArteryVascular Resistance

Substanzen

Carbon DioxideOxygen

Autor:innen und Institutionen

  • Department of Anesthesiology, Loyola University of Chicago, Maywood, IL 60153.

Zitieren

APA

Sami, H. M., Kleinman, B. S., & Lonchyna, V. A. (1991). Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry. Journal of Clinical Monitoring, 7(4), 309-312. https://doi.org/10.1007/bf01619351

Vancouver

Sami HM, Kleinman BS, Lonchyna VA. Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry. Journal of Clinical Monitoring. 1991;7(4):309-312. doi:10.1007/bf01619351

BibTeX
@article{Sami1991,
  author = {Sami, H M and Kleinman, B S and Lonchyna, V A},
  title = {Central venous pulsations associated with a falsely low oxygen saturation measured by pulse oximetry},
  journal = {Journal of Clinical Monitoring},
  year = {1991},
  volume = {7},
  number = {4},
  pages = {309--312},
  doi = {10.1007/bf01619351}
}