EUVIMEDEUVIMED entdecken
Beobachtungsstudie

Midline Catheters as an Alternative for Central Venous Catheters in Venous Pressure Monitoring: A Single Center Experience

Brendan Sweeney iD, Justin Kim iD, Ariana Adnani, Huma Saleem iD, Seda Akben iD, Yin Rong Alvina Teo, Eduard Shaykhinurov iD, Maria Wu, Rishika Bheem, Samita Islam iD, Danielle Davison iD, Katrina Hawkins iD und 3 weitere

Journal of Intensive Care Medicine · 2025

DOI: 10.1177/08850666251368867 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Journal of Intensive Care Medicine, 2025. Verfasst von Brendan Sweeney und 14 weiteren Autor:innen. Beteiligte Einrichtung: School of Medicine and Health Sciences, The George Washington University, Washington, DC, USA. Hauptthemen laut MEDLINE: Central Venous Catheters, Central Venous Pressure, Catheterization, Central Venous, Catheterization, Peripheral. Die Arbeit nennt 37 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Midline catheters (ML) are long peripheral intravenous catheters placed in an upper extremity above the antecubital fossa via the basilic, cephalic, or brachial veins. These provide safe and comfortable long-term vascular access for critically ill patients. Central venous pressures (CVP) are obtained from central venous catheters (CVC) and are often used as resuscitation parameters. Given the proximity of ML to the axillary vein, we investigated if midline venous pressure (MVP) is comparable to CVP.

Methods

This is an observational study conducted in adult critical care patients at a tertiary academic center. Inclusion criteria were patients with a CVC in the subclavian or internal jugular veins and a ML in place as part of standard ICU care. Pressure measurements were recorded from both catheters every 15 min over a 60-min period. Demographic, clinical, and physiological data points were collected. Continuous variables were analyzed using the t-test. Pearson correlation was used to evaluate the relationship between the paired variables.

Results

We enrolled 50 patients with 5 pressure measurements taken per patient (n = 250). The mean MVP and CVP were 10.6 ± 6.4 mm Hg and 9.1 ± 6.3 mm Hg, respectively ( P < .001). In Bland Altman analysis, the mean bias was −1.48 ± 3.99 mm Hg with limits of agreement of −9.3 mm Hg to 6.3 mm Hg. The Pearson correlation coefficient was 0.8 ( P < .001).

Conclusions

Our study investigated MVP as a correlate of CVP. Our results show a mean bias of −1.48 ± 3.99 mm Hg and a strong positive Pearson correlation coefficient of 0.8 between the MVP and the CVP. The large limits of agreement indicate MVP and CVP are not interchangeable. Despite this, clinically significant pressure values from CVCs and MLs trend similarly.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Central Venous CathetersCentral Venous PressureCatheterization, Central VenousCatheterization, PeripheralBlood Pressure DeterminationHumansMaleFemaleMiddle AgedJugular VeinsAgedAdultSubclavian VeinCritical CareIntensive Care UnitsMonitoring, PhysiologicCritical Illness

Stichwörter der Autor:innen

cardiac outputcentral venous catheterhemodynamic monitoringhemodynamicsqualityvolume status

Autor:innen und Institutionen

  • School of Medicine and Health Sciences, The George Washington University, Washington, DC, USA.
  • College of Health Professions, Virginia Commonwealth University, Richmond, VA, USA.
  • Department of Anesthesia and Critical Care Medicine, George Washington University, Washington, DC, USA.
  • Department of Pulmonary and Critical Care Medicine, Yale New Haven Hospital, New Haven, CT, USA.

ORCID-Profile: Brendan Sweeney, Justin Kim, Huma Saleem, Seda Akben, Eduard Shaykhinurov, Samita Islam, Danielle Davison, Katrina Hawkins, Daniel King, David P Yamane

Zitieren

APA

Sweeney, B., Kim, J., Adnani, A., Saleem, H., Akben, S., Teo, Y. R. A., Shaykhinurov, E., Wu, M., Bheem, R., Islam, S., Davison, D., Hawkins, K., King, D., Ivanovic, S., & Yamane, D. P. (2025). Midline Catheters as an Alternative for Central Venous Catheters in Venous Pressure Monitoring: A Single Center Experience. Journal of Intensive Care Medicine. https://doi.org/10.1177/08850666251368867

Vancouver

Sweeney B, Kim J, Adnani A, Saleem H, Akben S, Teo YRA, et al. Midline Catheters as an Alternative for Central Venous Catheters in Venous Pressure Monitoring: A Single Center Experience. Journal of Intensive Care Medicine. 2025. doi:10.1177/08850666251368867

BibTeX
@article{Sweeney2025,
  author = {Sweeney, Brendan and Kim, Justin and Adnani, Ariana and Saleem, Huma and Akben, Seda and Teo, Yin Rong Alvina and Shaykhinurov, Eduard and Wu, Maria and Bheem, Rishika and Islam, Samita and Davison, Danielle and Hawkins, Katrina and King, Daniel and Ivanovic, Sasa and Yamane, David P},
  title = {Midline Catheters as an Alternative for Central Venous Catheters in Venous Pressure Monitoring: A Single Center Experience},
  journal = {Journal of Intensive Care Medicine},
  year = {2025},
  doi = {10.1177/08850666251368867}
}
Kostenlos starten →