EUVIMEDObjevte EUVIMED
Crossref

Synchronized intermittent mandatory ventilation with volume guarantee and pressure support in neonates: Detailed analysis of ventilator parameters

Andras Balajthy, Gergely Balazs iD, Tamas Kovacs, Gusztav Belteki

Pediatric Pulmonology · 2023

DOI: 10.1002/ppul.26384 ↗

At a glance

Journal article in Pediatric Pulmonology, 2023. Written by Andras Balajthy and 3 co-authors. Institution: Department of Pediatrics, Division of Neonatology, Faculty of Medicine, University of Debrecen, Debrecen, Hungary. Major MEDLINE topics: Ventilators, Mechanical, Intermittent Positive-Pressure Ventilation. Cited 4 times so far (Crossref); 16 references.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Objective

To analyse the relationship between peak inflating pressure, expired tidal volume, respiratory rate, and inspiratory time of volume‐guaranteed ventilator inflations and pressure‐supported spontaneous breaths during synchronized intermittent positive pressure mode with volume guarantee and pressure support (SIMV‐VG‐PS) in neonates.

Methods

Ventilator parameters were downloaded every second from 16 babies ventilated with SIMV‐VG‐PS mode using Dräger Babylog VN500 ventilators over 137 days. Transcutaneous carbon dioxide (tcCO 2 ) data were also collected. Data were computationally analysed using Python. The average of each ventilator parameter was determined during each minute separately for ventilator inflations and for spontaneous breaths. These values were compared and their effect on tcCO 2 levels was also analysed.

Results

The relationship between the peak inflating pressure of the volume guaranteed inflations (PIP VG ) and pressure‐supported spontaneous breaths (PIP PS ) was highly variable. The PIP PS /PIP VG ratio differed significantly from the value (0.66) targeted by clinicians (group median: 0.80, range: 0.50–1.00). PIP PS frequently exceeded PIP VG . When PIP PS /PIP VG was >0.66, the expired tidal volume and the rate of the pressure‐supported spontaneous breaths were also significantly ( p < 0.0001) higher, but there was no difference in tcCO 2 levels. The flow‐cycled spontaneous breaths had significantly shorter inspiratory times than ventilator inflations.

Conclusions

During SIMV‐VG‐PS it is difficult to ensure a pressure support level proportionate to the inflating pressure of ventilator inflations and to achieve the stability of tidal volumes.

Subjects

MeSH terms (MEDLINE)

Ventilators, MechanicalIntermittent Positive-Pressure VentilationHumansInfant, NewbornRespiratory RateTidal VolumeRespiration, Artificial

Author keywords

mechanical ventilationneonatal intensive carepressure supportsynchronised intermittent mandatory ventilationvolume guarantee

Authors and institutions

  • Department of Pediatrics, Division of Neonatology, Faculty of Medicine, University of Debrecen, Debrecen, Hungary.
  • Neonatal Intensive Care Unit, The Rosie Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge, UK.

ORCID profiles: Gergely Balazs

Cite

APA

Balajthy, A., Balazs, G., Kovacs, T., & Belteki, G. (2023). Synchronized intermittent mandatory ventilation with volume guarantee and pressure support in neonates: Detailed analysis of ventilator parameters. Pediatric Pulmonology. https://doi.org/10.1002/ppul.26384

Vancouver

Balajthy A, Balazs G, Kovacs T, Belteki G. Synchronized intermittent mandatory ventilation with volume guarantee and pressure support in neonates: Detailed analysis of ventilator parameters. Pediatric Pulmonology. 2023. doi:10.1002/ppul.26384

BibTeX
@article{Balajthy2023,
  author = {Balajthy, Andras and Balazs, Gergely and Kovacs, Tamas and Belteki, Gusztav},
  title = {Synchronized intermittent mandatory ventilation with volume guarantee and pressure support in neonates: Detailed analysis of ventilator parameters},
  journal = {Pediatric Pulmonology},
  year = {2023},
  doi = {10.1002/ppul.26384}
}