EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte Studie

Assisted ventilation immediately after birth with self-inflating bag versus T-piece resuscitator in preterm infants

M Khan, D Bateman, R Sahni, T A Leone

Journal of Neonatal-Perinatal Medicine · 2023

DOI: 10.3233/npm-210728 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Journal of Neonatal-Perinatal Medicine, 2023. Verfasst von M Khan und 3 weiteren Autor:innen. Beteiligte Einrichtung: Maimonides Medical Center, Brooklyn, NY, USA. Hauptthemen laut MEDLINE: Respiration, Artificial, Infant, Premature. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 15 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: To compare proportions of target range tidal volumes achieved with the self-inflating bag vs. the T-piece in resuscitation of preterm newborns at delivery. STUDY DESIGN: This randomized controlled trial was conducted at a tertiary Children’s Hospital. 20 preterm infants≤32 weeks’ gestational age with no congenital anomalies who needed positive pressure ventilation after birth were enrolled. Positive pressure ventilation was provided with the self-inflating bag or T-piece resuscitator. The primary outcome was proportion of inflations within a target range of 4–8 ml/kg. Chi-square and logistical regression analyses were performed. RESULTS: In the self-inflating bag (SIB) group 29% of inflations (117/419) and in the T-Piece (TP) group 51% of inflations (300/590) delivered expiratory tidal volume (TVe) of 4–8 ml/kg (p < 65.001). In the SIB group 60% of all inflations (254/419), and in the TP group 35% of all inflations (204/590) delivered TVe < 4 ml/kg (p < 0.001). In the SIB group 11% of all inflations (48/419), and in the TP group, 15% of all inflations (86/590) delivered TVe > 8 ml/kg (p = 0.18). The OR of having expiratory tidal volume of 4–8 ml/kg using the T-piece was 1.8 (CI 1.1–3.1), p = 0.02. CONCLUSION: Manual inflations provided by the TP deliver expiratory tidal volumes in the range of 4–8 ml/kg more consistently than SIB.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Respiration, ArtificialInfant, PrematureInfantChildInfant, NewbornHumansResuscitationPositive-Pressure RespirationIntermittent Positive-Pressure VentilationTidal Volume

Stichwörter der Autor:innen

Delivery room (DR)expiratory tidal volume (TVe)peak inspiratory pressure (PIP)positive end expiratory pressure (PEEP)positive pressure ventilation (PPV)

Autor:innen und Institutionen

  • Maimonides Medical Center, Brooklyn, NY, USA.
  • New York Presbyterian Morgan Stanley Children's Hospital, New York, NY, USA.

Zitieren

APA

Khan, M., Bateman, D., Sahni, R., & Leone, T. A. (2023). Assisted ventilation immediately after birth with self-inflating bag versus T-piece resuscitator in preterm infants. Journal of Neonatal-Perinatal Medicine. https://doi.org/10.3233/npm-210728

Vancouver

Khan M, Bateman D, Sahni R, Leone TA. Assisted ventilation immediately after birth with self-inflating bag versus T-piece resuscitator in preterm infants. Journal of Neonatal-Perinatal Medicine. 2023. doi:10.3233/npm-210728

BibTeX
@article{Khan2023,
  author = {Khan, M and Bateman, D and Sahni, R and Leone, T A},
  title = {Assisted ventilation immediately after birth with self-inflating bag versus T-piece resuscitator in preterm infants},
  journal = {Journal of Neonatal-Perinatal Medicine},
  year = {2023},
  doi = {10.3233/npm-210728}
}