Identifying Slim Modiolar Electrode Tip Fold‐Over With Intracochlear Electrocochleography
Otolaryngology–Head and Neck Surgery · 2024
DOI: 10.1002/ohn.587 ↗At a glance
Journal article in Otolaryngology–Head and Neck Surgery, 2024. Written by Jordan J Varghese and 8 co-authors. Institution: Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA. Major MEDLINE topics: Cochlear Implants, Cochlear Implantation. Cited 5 times so far (Crossref); 39 references. The full text is freely available in PubMed Central. Funded by: NIDCD NIH HHS, NCATS NIH HHS.
Abstract (original language)
Objective
To evaluate the predictive value of intracochlear electrocochleography (ECochG) for identifying tip fold‐over during cochlear implantation (CI) using the slim modiolar electrode (SME) array. Study Design Prospective cohort study.
Setting
Tertiary referral center.
Methods
From July 2022 to June 2023, 142 patients, including adults and children, underwent intracochlear ECochG monitoring during and after SME placement. Tone‐bursts were presented from 250 Hz to 2 kHz at 108 to 114 dB HL. A fast Fourier transform (FFT) allowed for frequency‐specific evaluation of ECochG response. ECochG patterns during insertion and postinsertion were evaluated using sensitivity and specificity analysis to predict tip fold‐over. Intraoperative plain radiographs served as a reference standard.
Results
Fifteen tip fold‐over cases occurred (10.6%) with significant ECochG response (>2 µV). Sixty‐one cases without tip fold‐over occurred (43.0%) with significant ECochG response. All tip fold‐overs had both a nontonotopic postinsertion sweep and nonrobust active insertion pattern. No patients with robust insertion or tonotopic sweep patterns had tip fold‐over. Sensitivity of detecting tip fold‐over when having both nonrobust insertion and nontonotopic sweep patterns was 100% (95% confidence inteval [CI] 78.2%‐100%), specificity was 68.9% (95% CI 55.7%‐80.1%), and the overall accuracy was 72.0% (95% CI 60.5%‐81.7%).
Conclusion
Intracochlear ECochG monitoring during cochlear implantation with the SME can be a valuable tool for identifying properly positioned electrode arrays. In cases where ECochG patterns are nonrobust on insertion and nontonotopic for electrode sweeps, there may be a concern for tip fold‐over, and intraoperative imaging is necessary to confirm proper insertion.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
ORCID profiles: Jordan J Varghese
Funding
- NIDCD NIH HHS
- NCATS NIH HHS
Cite
APA
Varghese, J. J., Walia, A., Lefler, S. M., Ortmann, A. J., Shew, M. A., Durakovic, N., Wick, C. C., Herzog, J. A., & Buchman, C. A. (2024). Identifying Slim Modiolar Electrode Tip Fold‐Over With Intracochlear Electrocochleography. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1002/ohn.587
Vancouver
Varghese JJ, Walia A, Lefler SM, Ortmann AJ, Shew MA, Durakovic N, et al. Identifying Slim Modiolar Electrode Tip Fold‐Over With Intracochlear Electrocochleography. Otolaryngology–Head and Neck Surgery. 2024. doi:10.1002/ohn.587
BibTeX
@article{Varghese2024,
author = {Varghese, Jordan J and Walia, Amit and Lefler, Shannon M and Ortmann, Amanda J and Shew, Matthew A and Durakovic, Nedim and Wick, Cameron C and Herzog, Jacques A and Buchman, Craig A},
title = {Identifying Slim Modiolar Electrode Tip Fold‐Over With Intracochlear Electrocochleography},
journal = {Otolaryngology–Head and Neck Surgery},
year = {2024},
doi = {10.1002/ohn.587}
}
EUVIMED