Advanced Inversion Recovery Single Point Imaging for High Spatial Resolution and Quantitative Oxygen Imaging

Boris Epel iD, Irene Canavesi iD, Mrignayani Kotecha iD

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70538 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2026. Written by Boris Epel and 2 co-authors. Institution: Oxygen Measurement Core, O2M Technologies, LLC, Chicago, Illinois, USA. Major MEDLINE topics: Oxygen, Image Processing, Computer-Assisted, Fibrosarcoma. 36 references. Funded by: NCI NIH HHS, NIH HHS, National Institutes of Health.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Purpose

The partial pressure of oxygen (pO 2 ) in tissue is a tightly regulated physiological quantity. Hypoxia is a hallmark of many diseases, and changes in tissue pO 2 underline critical developmental, metabolic, and physiological processes. Electron paramagnetic resonance oxygen imaging (EPROI) has emerged as the most precise technique to image pO 2 in vitro and in vivo. EPROI uses linear relationship between the relaxation rates (R 1 or R 2 ) of the spin probe OXO71 and pO 2 .

Methods

The current gold standard for EPROI is inversion recovery electron spin echo (IRESE) sequence with radial k ‐space acquisition scheme, which is fast, provides high pO 2 resolution using T 1 imaging, but suffers from low spatial resolution, especially at higher oxygen concentrations. The other method, single‐point imaging (SPI), which uses Cartesian k ‐space acquisition scheme and free‐induction decay (FID) readout, provides higher spatial resolution but has lower signal‐to‐noise ratio (SNR) and lower pO 2 accuracy, especially for T 2 * imaging. In this work, we introduce advanced inversion‐recovery single‐point imaging (AIR‐SPI), which combines inversion‐recovery SPI (IR‐SPI) with multi‐point FID averaging in the spatial domain. We evaluate the performance of AIR‐SPI in phantoms and in vivo imaging of a fibrosarcoma (FSA) tumor in a mouse leg.

Results

We demonstrate 2–8‐fold SNR improvement and higher spatial resolution for AIR‐SPI compared to IR‐SPI and IRESE, especially for high oxygen concentrations. We also demonstrate higher pO 2 precision for AIR‐SPI.

Conclusions

AIR‐SPI offers improvements in image SNR, spatial, and oxygen resolution over all existing EPR oxygen imaging methods, without increasing imaging time. Importantly, it uses smaller RF power than echo‐based methods, which is crucial for scaling EPROI to human applications.

Subjects

MeSH terms (MEDLINE)

OxygenImage Processing, Computer-AssistedFibrosarcomaElectron Spin Resonance SpectroscopyPhantoms, ImagingSignal-To-Noise RatioAnimalsMiceMagnetic Resonance Imaging

Author keywords

electron paramagnetic resonanceelectron spin–lattice relaxation timefunctional imagingoxygen imagingsingle point imaging

Substances

Oxygen

Authors and institutions

  • Oxygen Measurement Core, O2M Technologies, LLC, Chicago, Illinois, USA.
  • Department of Radiation and Cellular Oncology, The University of Chicago, Chicago, Illinois, USA.

ORCID profiles: Boris Epel, Irene Canavesi, Mrignayani Kotecha

Funding

  • NCI NIH HHS
  • NIH HHS
  • National Institutes of Health

Cite

APA

Epel, B., Canavesi, I., & Kotecha, M. (2026). Advanced Inversion Recovery Single Point Imaging for High Spatial Resolution and Quantitative Oxygen Imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70538

Vancouver

Epel B, Canavesi I, Kotecha M. Advanced Inversion Recovery Single Point Imaging for High Spatial Resolution and Quantitative Oxygen Imaging. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70538

BibTeX
@article{Epel2026,
  author = {Epel, Boris and Canavesi, Irene and Kotecha, Mrignayani},
  title = {Advanced Inversion Recovery Single Point Imaging for High Spatial Resolution and Quantitative Oxygen Imaging},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70538}
}
Start for free →