In Vivo Measurement of Taurine With Optimized Double‐Quantum Filtering Magnetic Resonance Spectroscopy
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70511 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Yufan Zhou und 7 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory for Biomedical Engineering of Ministry of Education, College of Biomedical Engineering and Instrument Science, Zhejiang University, Ha…. Hauptthemen laut MEDLINE: Taurine. Die Arbeit nennt 48 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: National Natural Science Foundation of China, Zhejiang Provincial Natural Science Foundation of China, National Key Research and Development Program of China.
Abstract (Originalsprache)
Purpose
To enhance the accuracy and reproducibility for in vivo taurine quantification using the double‐quantum filtering (DQF) technique.
Methods
Simulations and phantom experiments were conducted at both 7 T and 3 T to determine the optimal parameter τ by calculating the taurine purity index among other neighboring metabolites. The frequency‐selective read‐pulse bandwidth and the dephasing/rephasing gradient moments were further adjusted to reduce residual water side band interference. Rat models of C6 gliomas were compared to normal brain for taurine level at 7 T. Metabolomics was conducted to validate the accuracy of taurine quantification. Human gray and white matter taurine differences were revealed at 3 T. Test–retest reliability was evaluated on humans.
Results
Simulations and phantom experiments identified optimal τ for maximizing taurine signal purity as 102 ms at 7 T and 92 ms at 3 T. The best combinations of parameters for minimal water interference were determined at 3 T. Optimized DQF sequences yielded lower CRLBs. In rat brains, taurine concentration measured with optimized DQF sequences exhibited strong correlation with metabolomics, outperforming both conventional MRS and nonoptimized DQF sequences. Only the optimized method revealed significant taurine concentration differences between gray and white matter. Test–retest demonstrated good reproducibility in gray matter (ICC = 0.82) for the optimized method.
Conclusion
The optimized DQF method enhances taurine detection accuracy and robustness in both animal and human studies and was for the first time applied in human brain taurine test at 3 T.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Key Laboratory for Biomedical Engineering of Ministry of Education, College of Biomedical Engineering and Instrument Science, Zhejiang University, Hangzhou, Zhejiang, China.
- Department of Endocrinology, School of Medicine, Sir Run Run Shaw Hospital, Zhejiang University, Hangzhou, Zhejiang, China.
- Institute of Medical Magnetic Resonance and Molecular Imaging Technology, East China Normal University, Shanghai, China.
- Department of Nuclear Medicine and PET Center, The Second Affiliated Hospital of Zhejiang University School of Medicine, Hangzhou, Zhejiang, China.
ORCID-Profile: Yefeng Yao, Min Wang
Förderung
- National Natural Science Foundation of China
- Zhejiang Provincial Natural Science Foundation of China
- National Key Research and Development Program of China
Zitieren
APA
Zhou, Y., Ye, C., Zhou, J., Cao, Y., Lin, D., Yao, Y., Zhang, H., & Wang, M. (2026). In Vivo Measurement of Taurine With Optimized Double‐Quantum Filtering Magnetic Resonance Spectroscopy. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70511
Vancouver
Zhou Y, Ye C, Zhou J, Cao Y, Lin D, Yao Y, et al. In Vivo Measurement of Taurine With Optimized Double‐Quantum Filtering Magnetic Resonance Spectroscopy. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70511
BibTeX
@article{Zhou2026,
author = {Zhou, Yufan and Ye, Chenting and Zhou, Jiaqiang and Cao, Yang and Lin, Dingyi and Yao, Yefeng and Zhang, Hong and Wang, Min},
title = {In Vivo Measurement of Taurine With Optimized Double‐Quantum Filtering Magnetic Resonance Spectroscopy},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70511}
}
EUVIMED