Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Adam M Wright und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Imaging Sciences, Indiana University School of Medicine, Indianapolis, Indiana, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Heart, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIA NIH HHS, NIGMS NIH HHS, National Institute of General Medical Sciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To validate a method to assess cardiac and respiratory brain pulsations in slowly sampled dynamic MR scans. Cardiac and respiratory pulsations are key drivers of neurofluid circulation. However, resolving their temporal dynamics requires fast imaging, which is not achievable in many dynamic MR acquisitions.

Methods

We systematically validated TRACC‐PHYSIO, a Time‐domain Resolution‐Aligned Cross‐Correlation framework for estimating physiological coupling and time delays in dynamic MRI. This method uses simultaneously recorded cardiac and respiratory waveforms as external references and applies time‐shifted cross‐correlation to estimate physiological coupling strength and relative pulse delay in dynamic MRI data. Two metrics are derived: Peak Coupling Coefficient (Peak CorrCoeff), which quantifies coupling strength, and TimeDelay, which estimates the relative arrival time of the physiological impulse in the brain. TRACC‐PHYSIO was first evaluated with in vivo fMRI (TR = 363 ms) by comparing the Peak CorrCoeff with physiological bandpower derived from spectral analysis. Performance was further assessed using simulations of realistically modeled dynamic MR signals across repetition time (TR = 50–3000 ms) and acquisition durations (60–360 s).

Results

In fMRI, TRACC‐derived cardiac and respiratory Peak CorrCoeff were strongly associated with their respective spectrum‐derived physiological bandpower (all tests Pearson r > 0.90). In simulations, both Peak CorrCoeff and TimeDelay were estimated with no mean bias and low temporal errors across TRs and acquisition durations.

Conclusion

TRACC‐PHYSIO is a validated time‐domain framework for quantifying cardiac and respiratory coupling strength and estimating millisecond‐scale relative pulse delays in standard dynamic MR acquisitions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingBrainHeartImage Processing, Computer-AssistedHumansAlgorithmsComputer SimulationReproducibility of ResultsRespiration

Stichwörter der Autor:innen

cardiac and respiratory pulsationscoupling strengthdynamic diffusion‐weighted imaging (dynDWI)functional MRI (fMRI)neurofluid dynamicspulse time delays

Autor:innen und Institutionen

  • Department of Radiology and Imaging Sciences, Indiana University School of Medicine, Indianapolis, Indiana, USA.
  • Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana, USA.

ORCID-Profile: Adam M Wright, Qiuting Wen

Förderung

  • NIA NIH HHS
  • NIGMS NIH HHS
  • National Institute of General Medical Sciences
  • National Institute on Aging

Zitieren

APA

Wright, A. M., Zhang, J., Tong, Y., & Wen, Q. (2026). TRACC ‐ PHYSIO : Time‐Domain Resolution‐Aligned Cross‐Correlation to Estimate PHYSIOlogical Coupling and Time Delays in Dynamic MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70446

Vancouver

Wright AM, Zhang J, Tong Y, Wen Q. TRACC ‐ PHYSIO : Time‐Domain Resolution‐Aligned Cross‐Correlation to Estimate PHYSIOlogical Coupling and Time Delays in Dynamic MRI. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70446

BibTeX
@article{Wright2026,
  author = {Wright, Adam M and Zhang, Jianing and Tong, Yunjie and Wen, Qiuting},
  title = {TRACC ‐ PHYSIO : Time‐Domain Resolution‐Aligned Cross‐Correlation to Estimate PHYSIOlogical Coupling and Time Delays in Dynamic MRI},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70446}
}
Kostenlos starten →