Evaluation of an interleaved acquisition scheme for improved robustness of channel‐wise relative B 1 + mapping at 7 T
Magnetic Resonance in Medicine · 2026-02 · 95 (2) · 1157-1168
DOI: 10.1002/mrm.70101 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 29.09.2025. Verfasst von Nico Egger und 9 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Heart, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The purpose was to evaluate, whether an interleaved acquisition scheme for a fast relative mapping method at 7 T reduces the likelihood of errors from exceeding the linear flip angle (FA) regime compared to a conventional sequential acquisition.
Methods
Simulations of a channel‐wise relative mapping sequence were performed for sequential and interleaved acquisition schemes at different reference voltages (≙ different FAs). The simulations were performed for a phantom, the heart and the prostate and were based on 7 T ground‐truth (GT) data acquired with an actual FA imaging sequence (phantom) or obtained from electromagnetic field simulations (heart/prostate). Acquisition schemes were evaluated based on their signal linearity by calculating a normalized mean FA error between simulated signal intensities and GT data. Additionally, validation measurements of relative maps with the sequential and interleaved acquisition schemes were acquired for the phantom.
Results
Validation measurements showed a good agreement with the simulation results for both acquisition schemes and displayed stronger deviations to the GT data for the sequential scheme. The quantitative evaluation yielded higher FA errors for the sequential acquisition scheme for all three regions and all simulated reference voltages. At the same level of error, mean signals were higher for the interleaved acquisition scheme in all cases. Differences between interleaved and sequential acquisition schemes were most pronounced in the steady‐state.
Conclusion
An interleaved acquisition of channel‐wise relative maps extends the range of the linear FA regime, reducing the likelihood of errors and increasing the robustness of the approach.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
- Electrical Engineering and Information Technology, University of Applied Sciences - FH Aachen, Aachen, Germany.
- Siemens Healthcare, Erlangen, Germany.
- Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.
- Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, Minnesota, USA.
- Division of Medical Physics in Radiology, German Cancer Research Centre (DKFZ), Heidelberg, Germany.
ORCID-Profile: Nico Egger, Laurent Ruck, Jürgen Herrler, Sebastian Schmitter, Armin Michael Nagel
Zitieren
APA
Egger, N., Ruck, L., Nagelstraßer, S., Schirmer, J., Wildenberg, S., Bitz, A., Herrler, J., Schmitter, S., Uder, M., & Nagel, A. M. (2026). Evaluation of an interleaved acquisition scheme for improved robustness of channel‐wise relative B 1 + mapping at 7 T. Magnetic Resonance in Medicine, 95(2), 1157-1168. https://doi.org/10.1002/mrm.70101
Vancouver
Egger N, Ruck L, Nagelstraßer S, Schirmer J, Wildenberg S, Bitz A, et al. Evaluation of an interleaved acquisition scheme for improved robustness of channel‐wise relative B 1 + mapping at 7 T. Magnetic Resonance in Medicine. 2026;95(2):1157-1168. doi:10.1002/mrm.70101
BibTeX
@article{Egger2026,
author = {Egger, Nico and Ruck, Laurent and Nagelstraßer, Sophia and Schirmer, Judith and Wildenberg, Saskia and Bitz, Andreas and Herrler, Jürgen and Schmitter, Sebastian and Uder, Michael and Nagel, Armin Michael},
title = {Evaluation of an interleaved acquisition scheme for improved robustness of channel‐wise relative B 1 + mapping at 7 T},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {95},
number = {2},
pages = {1157--1168},
doi = {10.1002/mrm.70101}
}
EUVIMED