EUVIMEDEUVIMED entdecken
Crossref

High Spatial‐Resolution and Acquisition‐Efficiency Cardiac MR T1 Mapping Based on Radial bSSFP and a Low‐Rank Tensor Constraint

Juan Gao iD, Yiwen Gong, Yixin Emu, Zhuo Chen, Haiyang Chen iD, Fan Yang, Zekang Ding, Sha Hua, Wei Jin, Chenxi Hu iD

Journal of Magnetic Resonance Imaging · 2025

DOI: 10.1002/jmri.29564 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, 2025. Verfasst von Juan Gao und 9 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China. Hauptthemen laut MEDLINE: Heart, Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Cardiomyopathies. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 40 Literaturangaben. Gefördert von: Shanghai Science and Technology Commission, Shanghai Oriental Talents Youth Project, Huangpu District Health System Talent Training Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Cardiac T1 mapping is valuable for evaluating myocardial fibrosis, yet its resolution and acquisition efficiency are limited, potentially obscuring visualization of small pathologies.

Purpose

To develop a technique for high‐resolution cardiac T1 mapping with a less‐than‐100‐millisecond acquisition window based on radial MOdified Look‐Locker Inversion recovery (MOLLI) and a calibrationless space‐contrast‐coil locally low‐rank tensor (SCC‐LLRT) constrained reconstruction. Study Type Prospective. Subjects/Phantom Sixteen healthy subjects (age 25 ± 3 years, 44% females) and 12 patients with suspected cardiomyopathy (age 57 ± 15 years, 42% females), NiCl2‐agar phantom. Field Strength/Sequence 3‐T, standard MOLLI, radial MOLLI, inversion‐recovery spin‐echo, late gadolinium enhancement. Assessment SCC‐LLRT was compared to a conventional locally low‐rank (LLR) method through simulations using Normalized Root‐Mean‐Square Error (NRMSE) and Structural Similarity Index Measure (SSIM). Radial MOLLI was compared to standard MOLLI across phantom, healthy subjects, and patients. Three independent readers subjectively evaluated the quality of T1 maps using a 5‐point scale (5 = best). Statistical Tests Paired t ‐test, Wilcoxon signed‐rank test, intraclass correlation coefficient analysis, linear regression, Bland–Altman analysis. P < 0.05 was considered statistically significant.

Results

In simulations, SCC‐LLRT demonstrated a significant improvement in NRMSE and SSIM compared to LLR. In phantom, both radial MOLLI and standard MOLLI provided consistent T1 estimates across different heart rates. In healthy subjects, radial MOLLI exhibited a significantly lower mean T1 (1115 ± 39 msec vs. 1155 ± 36 msec), similar T1 SD (74 ± 14 msec vs. 67 ± 23 msec, P = 0.20), and similar T1 reproducibility (28 ± 18 msec vs. 22 ± 15 msec, P = 0.34) compared to standard MOLLI. In patients, the proposed method significantly improved the sharpness of myocardial boundaries (4.50 ± 0.65 vs. 3.25 ± 0.43), the conspicuity of papillary muscles and fine structures (4.33 ± 0.74 vs. 3.33 ± 0.47), and artifacts (4.75 ± 0.43 vs. 3.83 ± 0.55). The reconstruction time for a single slice was 5.2 hours. Data Conclusion The proposed method enables high‐resolution cardiac T1 mapping with a short acquisition window and improved image quality. Evidence Level 1 Technical Efficacy Stage 1

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HeartMagnetic Resonance ImagingImage Processing, Computer-AssistedCardiomyopathiesHumansFemaleMiddle AgedPhantoms, ImagingMaleAdultProspective StudiesReproducibility of ResultsAlgorithmsAgedImage Interpretation, Computer-AssistedContrast MediaYoung AdultComputer SimulationFibrosis

Stichwörter der Autor:innen

accelerationcardiac T1 mappinghigh resolutionlow‐rankradialtensor

Substanzen

Contrast Media

Autor:innen und Institutionen

  • School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
  • Department of Cardiovascular Medicine, Heart Failure Center, Ruijin Hospital and Ruijin Hospital Lu Wan Branch, Shanghai Jiao Tong University School of Medicine, Shanghai, China.

ORCID-Profile: Juan Gao, Haiyang Chen, Chenxi Hu

Förderung

  • Shanghai Science and Technology Commission
  • Shanghai Oriental Talents Youth Project
  • Huangpu District Health System Talent Training Project
  • National Natural Science Foundation of China

Zitieren

APA

Gao, J., Gong, Y., Emu, Y., Chen, Z., Chen, H., Yang, F., Ding, Z., Hua, S., Jin, W., & Hu, C. (2025). High Spatial‐Resolution and Acquisition‐Efficiency Cardiac MR T1 Mapping Based on Radial bSSFP and a Low‐Rank Tensor Constraint. Journal of Magnetic Resonance Imaging. https://doi.org/10.1002/jmri.29564

Vancouver

Gao J, Gong Y, Emu Y, Chen Z, Chen H, Yang F, et al. High Spatial‐Resolution and Acquisition‐Efficiency Cardiac MR T1 Mapping Based on Radial bSSFP and a Low‐Rank Tensor Constraint. Journal of Magnetic Resonance Imaging. 2025. doi:10.1002/jmri.29564

BibTeX
@article{Gao2025,
  author = {Gao, Juan and Gong, Yiwen and Emu, Yixin and Chen, Zhuo and Chen, Haiyang and Yang, Fan and Ding, Zekang and Hua, Sha and Jin, Wei and Hu, Chenxi},
  title = {High Spatial‐Resolution and Acquisition‐Efficiency Cardiac MR T1 Mapping Based on Radial bSSFP and a Low‐Rank Tensor Constraint},
  journal = {Journal of Magnetic Resonance Imaging},
  year = {2025},
  doi = {10.1002/jmri.29564}
}