SelExNet : A Self‐Supervised Physics‐Informed Framework for Multi‐Channel Joint RF and Gradient Waveform Optimization in 2D Spatially Selective Excitation
Magnetic Resonance in Medicine · 2026-09 · 96 (3) · 1219-1234
DOI: 10.1002/mrm.70431 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 14.05.2026. Verfasst von Yuliang Xiao und 5 weiteren Autor:innen. Beteiligte Einrichtung: Physical Sciences Platform, Sunnybrook Research Institute, Toronto, Ontario, Canada. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Brain. Die Arbeit nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada, CIHR, Canadian Institutes of Health Research.
Abstract (Originalsprache)
Purpose
To introduce SelExNet: a self‐supervised framework for two‐dimensional spatially selective excitation that jointly optimizes radiofrequency (RF) pulses and gradient waveforms, and extends to multi‐channel transmission MRI.
Methods
Building on prior RF‐only and joint RF‐gradient optimization approaches, SelExNet couples neural RF and gradient generators with a differentiable Bloch simulator to enable self‐supervised pulse optimization without requiring pre‐designed target pulses. The framework jointly designs RF pulses and parameterized variable‐density spiral gradient waveforms for both single‐ and multi‐channel transmission, with patient‐specific adaptation using measured, previously unseen and maps.
Results
Joint optimization of RF and gradients improved excitation fidelity compared to RF‐only optimization. In phantom experiments with synthetic field maps, pretrained pulses showed distortions, whereas fine‐tuned pulses restored geometry and uniformity. In vivo studies demonstrated anatomically precise excitation, with fine‐tuning improving sharpness and reducing off‐target signal.
Conclusion
The proposed framework enables joint RF‐gradient design and extends self‐supervised pulse optimization to multi‐channel transmission MRI. SelExNet achieves high‐fidelity, anatomically precise excitation and demonstrates robustness to field inhomogeneities, offering a scalable pathway for ultra‐high field imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Physical Sciences Platform, Sunnybrook Research Institute, Toronto, Ontario, Canada.
- Department of Medical Biophysics, Temerty Faculty of Medicine, University of Toronto, Toronto, Ontario, Canada.
- Siemens Healthcare Limited, Oakville, Ontario, Canada.
ORCID-Profile: Yuliang Xiao, Jason Rock, Zhe Wu, Jamie Near, Mark Chiew, Simon J Graham
Förderung
- Natural Sciences and Engineering Research Council of Canada
- CIHR
- Canadian Institutes of Health Research
Zitieren
APA
Xiao, Y., Rock, J., Wu, Z., Near, J., Chiew, M., & Graham, S. J. (2026). SelExNet : A Self‐Supervised Physics‐Informed Framework for Multi‐Channel Joint RF and Gradient Waveform Optimization in 2D Spatially Selective Excitation. Magnetic Resonance in Medicine, 96(3), 1219-1234. https://doi.org/10.1002/mrm.70431
Vancouver
Xiao Y, Rock J, Wu Z, Near J, Chiew M, Graham SJ. SelExNet : A Self‐Supervised Physics‐Informed Framework for Multi‐Channel Joint RF and Gradient Waveform Optimization in 2D Spatially Selective Excitation. Magnetic Resonance in Medicine. 2026;96(3):1219-1234. doi:10.1002/mrm.70431
BibTeX
@article{Xiao2026,
author = {Xiao, Yuliang and Rock, Jason and Wu, Zhe and Near, Jamie and Chiew, Mark and Graham, Simon J},
title = {SelExNet : A Self‐Supervised Physics‐Informed Framework for Multi‐Channel Joint RF and Gradient Waveform Optimization in 2D Spatially Selective Excitation},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {3},
pages = {1219--1234},
doi = {10.1002/mrm.70431}
}