EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI

Hongbae Jeong iD, Andrew Smith, Ananda Kumar

Magnetic Resonance in Medicine · 2026-08 · 96 (2) · 883-891

DOI: 10.1002/mrm.70389 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, published online on 2026-04-26. Written by Hongbae Jeong and 2 co-authors. Institution: Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Marylan…. Major MEDLINE topics: Magnetic Resonance Imaging, Pacemaker, Artificial, Defibrillators, Implantable. Cited 1 times so far (Crossref); 32 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

Patients with cardiac implantable electronic devices (CIEDs) may receive replacement device components over time, resulting in mixed‐manufacturer systems that pose unknown, potentially higher risk under MRI exposure. It is not clear how to efficiently evaluate whether adding a non‐original implantable pulse generator (IPG) changes the system's radiofrequency (RF)–induced heating response. This study aimed to develop a least‐burdensome RF safety evaluation method for MR Conditional labeling of CIEDs. Critical parameters were characterized and demonstrated as an IPG equivalency test method that supports leveraging existing single‐manufacturer MR‐Conditional labeling for mixed‐manufacturer systems.

Methods

The transfer function model of the pacing lead was measured using the original system (Lead A + IPG A ) or a third‐party IPGs (Lead A + IPG B or IPG C ). Deposited power near the lead tip was compared for 50 Ω and open‐end conditions, and lead or IPG impedances were measured at 64 and 128 MHz.

Results

Lead tip heating was not significantly different between original and third‐party IPGs configurations ( p < 0.001). RF‐induced heating responses were outside of measurement uncertainty when the lead was connected with a 50 Ω resistor and some open‐end conditions. The results showed that small changes in IPG impedance (< 2.5 Ω) to the lead‐IPG system impedance (> 80.2 Ω) may not significantly alter RF‐induced heating, remaining within measurement uncertainties.

Conclusion

A simplified IPG impedance equivalency test has been developed to facilitate MRI access for patients with CIEDs that undergo IPG interchange. This test aids assessment of whether replacing an original IPG model with a different model may significantly increase RF‐induced heating compared to the original system.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance ImagingPacemaker, ArtificialDefibrillators, ImplantableElectric ImpedanceEquipment DesignHumansRadio Waves

Author keywords

CIEDMR safetyRF‐induced heatingactive implantable devicesmixed‐manufacturer systempulse generator

Authors and institutions

  • Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Maryland, USA.
  • Office of Cardiovascular Devices, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Maryland, USA.

ORCID profiles: Hongbae Jeong

Cite

APA

Jeong, H., Smith, A., & Kumar, A. (2026). An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI. Magnetic Resonance in Medicine, 96(2), 883-891. https://doi.org/10.1002/mrm.70389

Vancouver

Jeong H, Smith A, Kumar A. An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI. Magnetic Resonance in Medicine. 2026;96(2):883-891. doi:10.1002/mrm.70389

BibTeX
@article{Jeong2026,
  author = {Jeong, Hongbae and Smith, Andrew and Kumar, Ananda},
  title = {An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {96},
  number = {2},
  pages = {883--891},
  doi = {10.1002/mrm.70389}
}