Consistent Quadratic Phase Formation in 3D Fast Spin Echo Using Frequency‐Modulated RF Pulses
Magnetic Resonance in Medicine · 2026-08 · 96 (2) · 666-681
DOI: 10.1002/mrm.70368 ↗At a glance
Journal article in Magnetic Resonance in Medicine, published online on 2026-04-02. Written by Naoharu Kobayashi and 1 co-authors. Institution: Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA. Major MEDLINE topics: Imaging, Three-Dimensional, Magnetic Resonance Imaging, Brain. 43 references. The full text is freely available (open access, CC BY 4.0). Funded by: NIBIB NIH HHS, NIH HHS, National Institutes of Health.
Abstract (original language)
Purpose
Frequency‐modulated (FM) RF pulses achieve broadband excitation with low RF peak power, which is required in MRI with inhomogeneous magnetic fields. However, the quadratic phase generated with FM pulses makes it difficult to use them in fast spin echo (FSE), because even and odd refocused echoes have different spatial phase profiles. In this article, we formulate the condition that can generate a consistent quadratic phase in the two echo components in 3D FSE using FM pulses.
Methods
The consistent quadratic phase condition in 3D FM‐FSE was formulated with the Cayley‐Klein parameters. The B 1 + ‐dependent phase in FM excitation was compensated by adjusting the initial phase of FM pulses. The consistent quadratic phase in 3D FM‐FSE and the impacts from the B 1 + ‐dependent phase were simulated with extended phase graph (EPG). The EPG simulation was validated by comparing with experimental results at 3 T. Finally, in vivo brain imaging was conducted with T 1 ‐ and T 2 ‐weighted contrasts.
Results
In EPG simulation, a consistent quadratic phase was generated in all echoes in 3D FM‐FSE under the formulated condition. The EPG simulation matched well with experimental results. The B 1 + ‐dependent phase adjustment on FM pulses improved the magnitude profile of the refocused echoes. Despite the presence of large field inhomogeneity, 3D FM‐FSE achieved in vivo brain imaging with good T 1 ‐ and T 2 ‐contrasts.
Conclusion
Achieving consistent quadratic phase in all refocused echoes in 3D FM‐FSE makes it possible to use FM pulses in the non‐adiabatic regime. 3D FM‐FSE is a promising pulse sequence for MRI in inhomogeneous magnetic fields.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
Funding
- NIBIB NIH HHS
- NIH HHS
- National Institutes of Health
Cite
APA
Kobayashi, N., & Garwood, M. (2026). Consistent Quadratic Phase Formation in 3D Fast Spin Echo Using Frequency‐Modulated RF Pulses. Magnetic Resonance in Medicine, 96(2), 666-681. https://doi.org/10.1002/mrm.70368
Vancouver
Kobayashi N, Garwood M. Consistent Quadratic Phase Formation in 3D Fast Spin Echo Using Frequency‐Modulated RF Pulses. Magnetic Resonance in Medicine. 2026;96(2):666-681. doi:10.1002/mrm.70368
BibTeX
@article{Kobayashi2026,
author = {Kobayashi, Naoharu and Garwood, Michael},
title = {Consistent Quadratic Phase Formation in 3D Fast Spin Echo Using Frequency‐Modulated RF Pulses},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {2},
pages = {666--681},
doi = {10.1002/mrm.70368}
}