A Hybrid RF Coil for Whole‐Brain Imaging in Non‐Human Primates at 7 T
Magnetic Resonance in Medicine · 2026-08 · 96 (2) · 996-1010
DOI: 10.1002/mrm.70361 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 24.03.2026. Verfasst von Elias Djaballah und 5 weiteren Autor:innen. Beteiligte Einrichtung: Cognitive Neuroimaging Unit, INSERM, CEA, Université Paris-Saclay, NeuroSpin Center, GIF-SUR-YVETTE, France. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Brain Mapping. Die Arbeit nennt 71 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The French Alternative Energies and Atomic Energy Commission, Agence Nationale de la Recherche, Fondation Leducq.
Abstract (Originalsprache)
Purpose
Ultra‐high field MRI (≥ 7 T) offers unprecedented potential for mapping non‐human primate (NHP) brain function. However, complex electromagnetic systems are required to meet the challenges of UHF imaging and the mechanical demands of awake NHP studies. To address these challenges, we developed a hybrid RF coil system optimized for whole‐brain fMRI of macaques at 7 T.
Methods
The hybrid coil system combines a 6‐channel preamplifier‐decoupled transceive dipole array with a 16‐channel loop receive array, housed in a compact structure designed for awake imaging that maintains open visual fields. Key design features include compatibility with sphinx‐position monkey chairs and head fixation systems required for awake experiments. Electromagnetic simulations guided dipole design to optimize uniform transmit performance and deep penetration. Phantom and in vivo experiments validated these predictions using anesthetized macaques and evaluated the system's readiness for awake imaging.
Results
The hybrid coil demonstrated uniform distribution with phase‐shimmed transmission across the brain. It supported robust parallel imaging, enabling up to R = 3 × 2 acceleration factors for high‐resolution acquisitions. The achieved whole‐brain tSNR in fMRI is comparable to that of an existing 8‐TX/24‐RX coil designed for imaging anesthetized macaques. Critically, submillimeter (0.75 mm isotropic) resting‐state fMRI revealed clear default‐mode network connectivity, confirming the system's capability for high‐quality functional imaging.
Conclusion
By effectively addressing both the technical challenges of ultra‐high field MRI and the mechanical constraints associated with visual stimulation in awake NHPs, this hybrid coil system provides a powerful tool for advancing our understanding of primate brain function.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cognitive Neuroimaging Unit, INSERM, CEA, Université Paris-Saclay, NeuroSpin Center, GIF-SUR-YVETTE, France.
- BAOBAB, Université Paris-Saclay, CEA/Joliot/NeuroSpin, GIF-SUR-YVETTE, France.
- Université Paris-Saclay, CEA/DRF/IRFU/DACM, GIF-SUR-YVETTE, France.
ORCID-Profile: Paul-François Gapais, Michel Luong, Alexis Amadon, Qi Zhu
Förderung
- The French Alternative Energies and Atomic Energy Commission
- Agence Nationale de la Recherche
- Fondation Leducq
- Commissariat à l'Énergie Atomique et aux Énergies Alternatives
Zitieren
APA
Djaballah, E., Giacomini, É., Gapais, P. F., Luong, M., Amadon, A., & Zhu, Q. (2026). A Hybrid RF Coil for Whole‐Brain Imaging in Non‐Human Primates at 7 T. Magnetic Resonance in Medicine, 96(2), 996-1010. https://doi.org/10.1002/mrm.70361
Vancouver
Djaballah E, Giacomini É, Gapais PF, Luong M, Amadon A, Zhu Q. A Hybrid RF Coil for Whole‐Brain Imaging in Non‐Human Primates at 7 T. Magnetic Resonance in Medicine. 2026;96(2):996-1010. doi:10.1002/mrm.70361
BibTeX
@article{Djaballah2026,
author = {Djaballah, Elias and Giacomini, Éric and Gapais, Paul-François and Luong, Michel and Amadon, Alexis and Zhu, Qi},
title = {A Hybrid RF Coil for Whole‐Brain Imaging in Non‐Human Primates at 7 T},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {2},
pages = {996--1010},
doi = {10.1002/mrm.70361}
}
EUVIMED