Chemical Shift Separated and Compensated Ultra‐Short Echo‐Time Imaging

Martin Krämer iD, Lumeng Cui, Jürgen R Reichenbach iD, Stefan Sommer iD

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70327 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2026. Written by Martin Krämer and 3 co-authors. Institution: Institute of Diagnostic and Interventional Radiology, Jena University Hospital, Friedrich Schiller University Jena, Jena, Germany. Major MEDLINE topics: Image Enhancement, Adipose Tissue, Image Interpretation, Computer-Assisted, Magnetic Resonance Imaging. Cited 1 times so far (Crossref); 27 references. The full text is freely available (open access, CC BY 4.0). Funded by: Deutsche Forschungsgemeinschaft.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

To develop a modified binomial excitation scheme for ultra‐short echo‐time (UTE) imaging that allows for the separation of water and fat signals and the correction of chemical shift artifacts.

Methods

Theoretical derivation and numerical Bloch simulations were performed for a modified water‐excitation binomial pulse scheme to calculate fat and water signals as well as a chemical shift corrected UTE image from two acquisitions with these configurations. Chemical shift correction was applied at the readout level by multiplying the fat signal with a corresponding linear phase term. Measurements with the proposed binomial pulse scheme were performed at 3 T and 7 T.

Results

Imaging results at 7 T and 3 T showed robust fat and water separation and images free of chemical shift‐induced blurring after applying the proposed method. Compared to standard UTE, the corrected UTE images exhibited improved tissue boundary delineation and improved visibility of fine details. Bloch simulations demonstrated that deviations of the phase differences between water and fat from the analytical solution were below 1% for typical UTE acquisition parameters.

Conclusion

The proposed binomial excitation scheme and post‐processing effectively separate water and fat signals and correct for chemical shift artifacts in UTE imaging, providing improved image quality, especially at high field strengths.

Subjects

MeSH terms (MEDLINE)

Image EnhancementAdipose TissueImage Interpretation, Computer-AssistedMagnetic Resonance ImagingImage Processing, Computer-AssistedAlgorithmsArtifactsReproducibility of ResultsHumansSensitivity and SpecificityComputer SimulationPhantoms, ImagingBody Water

Author keywords

DixonUTEbinomial pulseschemical shiftfat‐water separation

Authors and institutions

  • Institute of Diagnostic and Interventional Radiology, Jena University Hospital, Friedrich Schiller University Jena, Jena, Germany.
  • Medical Physics Group, Institute of Diagnostic and Interventional Radiology, Jena University Hospital, Friedrich Schiller University Jena, Jena, Germany.
  • Siemens Healthcare Limited, Burnaby, British Columbia, Canada.
  • Swiss Center for Musculoskeletal Imaging (SCMI), Balgrist Campus, Zurich, Switzerland.
  • Swiss Innovation Hub, Siemens Healthineers International AG, Zurich, Switzerland.

ORCID profiles: Martin Krämer, Jürgen R Reichenbach, Stefan Sommer

Funding

  • Deutsche Forschungsgemeinschaft

Cite

APA

Krämer, M., Cui, L., Reichenbach, J. R., & Sommer, S. (2026). Chemical Shift Separated and Compensated Ultra‐Short Echo‐Time Imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70327

Vancouver

Krämer M, Cui L, Reichenbach JR, Sommer S. Chemical Shift Separated and Compensated Ultra‐Short Echo‐Time Imaging. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70327

BibTeX
@article{Krmer2026,
  author = {Krämer, Martin and Cui, Lumeng and Reichenbach, Jürgen R and Sommer, Stefan},
  title = {Chemical Shift Separated and Compensated Ultra‐Short Echo‐Time Imaging},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70327}
}
Start for free →