K‐ CC ‐ MoCo : A Fast k ‐Space‐Based Respiratory Motion Correction for Highly Accelerated First‐Pass Perfusion Cardiovascular MR
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70287 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Elisa Moya-Sáez und 10 weiteren Autor:innen. Beteiligte Einrichtung: ETSI de Telecomunicación, Universidad de Valladolid, Valladolid, Spain. Hauptthemen laut MEDLINE: Image Processing, Computer-Assisted, Heart, Coronary Artery Disease. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: LARSyS FCT, Fundação para a Ciência e a Tecnologia, MICIU/AEI and FSE+.
Abstract (Originalsprache)
Purpose
First‐pass perfusion cardiovascular MR (FPP‐CMR) enables the non‐invasive diagnosis of microcirculation and coronary artery disease. In free‐breathing FPP‐CMR, motion correction is usually performed in the image domain, requiring an initial reconstruction. This fact hinders its use in model‐based and deep learning reconstructions, which present remarkable performance in obtaining high‐quality images from highly accelerated acquisitions. We address this challenge by estimating and correcting respiratory motion in free‐breathing FPP‐CMR directly in k ‐space.
Methods
We propose K‐CC‐MoCo, an inter‐frame rigid motion correction approach formulated exclusively in k ‐space that handles dynamic contrast through a specifically targeted design of the normalized cross‐correlation (CC) objective function to deal with the dynamic contrast. In addition, an ROI‐based coil‐compression approach was employed to focus the optimization on the heart region. The proposed method was compared to state‐of‐the‐art image‐based registration using a digital phantom and real free‐breathing acquisitions with different accelerations.
Results
The proposed k ‐space‐based method is approximately 2× faster and can correct respiratory motion even at high acceleration factors (up to 50×), where the image‐based method fails due to severe undersampling artifacts. Notably, after K‐CC‐MoCo, the time‐averaged images are visibly less blurred. Quantitative metrics (SSIM, etc.) support this conclusion.
Conclusion
K‐CC‐MoCo outperforms image‐based correction in free‐breathing FPP‐CMR acquisitions accelerated up to 50×. Respiratory motion is estimated and corrected in k ‐space, enabling its use for model‐based and/or deep learning reconstructions from highly accelerated scans.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- ETSI de Telecomunicación, Universidad de Valladolid, Valladolid, Spain.
- Instituto de Investigación Biosanitaria de Valladolid (IBioVALL), Valladolid, Spain.
- Universidad Politécnica de Cartagena Member of European University of Technology EUT+, Cartagena, Spain.
- Philips Healthcare Iberia, Madrid, Spain.
- Center of Marine Sciences-CCMAR, Faro, Portugal.
- Institute for Systems and Robotics - Lisboa and Department of Bioengineering, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.
ORCID-Profile: Elisa Moya-Sáez, Rosa-María Menchón-Lara, Javier Sánchez-González, Andreia S Gaspar, Carlos Real, Carlos Galán-Arriola, Rita G Nunes, Borja Ibanez, Teresa M Correia, Carlos Alberola-López
Förderung
- LARSyS FCT
- Fundação para a Ciência e a Tecnologia
- MICIU/AEI and FSE+
- Ministerio de Ciencia e Innovación of Spain
- 'la Caixa' Foundation
Zitieren
APA
Moya-Sáez, E., Menchón-Lara, R. M., Sánchez-González, J., Carvalho, C. N., Gaspar, A. S., Real, C., Galán-Arriola, C., Nunes, R. G., Ibanez, B., Correia, T. M., & Alberola-López, C. (2026). K‐ CC ‐ MoCo : A Fast k ‐Space‐Based Respiratory Motion Correction for Highly Accelerated First‐Pass Perfusion Cardiovascular MR. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70287
Vancouver
Moya-Sáez E, Menchón-Lara RM, Sánchez-González J, Carvalho CN, Gaspar AS, Real C, et al. K‐ CC ‐ MoCo : A Fast k ‐Space‐Based Respiratory Motion Correction for Highly Accelerated First‐Pass Perfusion Cardiovascular MR. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70287
BibTeX
@article{MoyaSez2026,
author = {Moya-Sáez, Elisa and Menchón-Lara, Rosa-María and Sánchez-González, Javier and Carvalho, Catarina N and Gaspar, Andreia S and Real, Carlos and Galán-Arriola, Carlos and Nunes, Rita G and Ibanez, Borja and Correia, Teresa M and Alberola-López, Carlos},
title = {K‐ CC ‐ MoCo : A Fast k ‐Space‐Based Respiratory Motion Correction for Highly Accelerated First‐Pass Perfusion Cardiovascular MR},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70287}
}
EUVIMED