Prospective Respiratory Motion Correction Using Machine Learning and Pilot Tone ( PROMPT ) in Cardiac MRI
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70149 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Yue Pan und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, The Ohio State University, Columbus, Ohio, USA. Hauptthemen laut MEDLINE: Heart, Machine Learning, Image Processing, Computer-Assisted, Magnetic Resonance Imaging. Die Arbeit nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NHLBI NIH HHS, Robert F. Wolfe and Edgar T. Wolfe Foundation, National Heart, Lung, and Blood Institute.
Abstract (Originalsprache)
Purpose
To implement and evaluate the feasibility of a Pilot Tone (PT)‐based prospective gating and tracking technique, which uses a long short‐term memory (LSTM) neural network to predict respiratory motion from PT signals.
Methods
A subject‐specific calibration scan consisting of 100 ECG‐triggered single‐shot images was performed. Respiratory motion was estimated from the images and PT data were processed to extract the respiratory component. The LSTM model was trained to predict respiratory motion from PT signals. During respiratory‐corrected scans, PROMPT‐predicted slice‐shifting parameters were used to update gating information and slice position for each heartbeat before image acquisition. The method was retrospectively evaluated in 12 healthy volunteers, comparing LSTM with linear and polynomial regression models using normalized root mean square error in decibels (NRMSE dB ) and mean absolute error (MAE). PROMPT was then implemented in late gadolinium enhancement (LGE) and compared with free‐breathing retrospective gating in 14 patients. Residual in‐plane motion was calculated to assess performance.
Results
The LSTM model achieved an NRMSE dB of −7.20 dB and an MAE of 1.97 mm between predicted and actual motion, demonstrating significantly higher accuracy than either regression models ( p < 0.05). The residual in‐plane motion in PROMPT‐LGE was significantly lower than in FB‐LGE (1.22 ± 0.38 mm vs. 1.35 ± 0.48 mm, p = 0.033).
Conclusion
The proposed respiratory motion correction approach was successfully implemented. The LSTM‐based predictive model outperformed linear and polynomial regression models. In LGE imaging, PROMPT significantly reduced in‐plane motion and showed potential for limiting through‐plane motion compared to the clinical protocol.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, The Ohio State University, Columbus, Ohio, USA.
- Dorothy M. Davis Heart and Lung Research Institute, The Ohio State University Wexner Medical Center, Columbus, Ohio, USA.
- MR R&D, Siemens Medical Solutions USA, Inc., Malvern, Pennsylvania, USA.
- Siemens Healthineers AG, Forchheim, Germany.
- Department of Diagnostic and Interventional Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.
- Center for Biostatistics, Department of Biomedical Informatics, The Ohio State University College of Medicine, Columbus, Ohio, USA.
ORCID-Profile: Yue Pan, Kelvin Chow, Mario Bacher, Rizwan Ahmad, Orlando P Simonetti
Förderung
- NHLBI NIH HHS
- Robert F. Wolfe and Edgar T. Wolfe Foundation
- National Heart, Lung, and Blood Institute
Zitieren
APA
Pan, Y., Jin, N., Chow, K., Bacher, M., Speier, P., Yildiz, V. O., Tong, M., Han, Y., Ahmad, R., & Simonetti, O. P. (2026). Prospective Respiratory Motion Correction Using Machine Learning and Pilot Tone ( PROMPT ) in Cardiac MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70149
Vancouver
Pan Y, Jin N, Chow K, Bacher M, Speier P, Yildiz VO, et al. Prospective Respiratory Motion Correction Using Machine Learning and Pilot Tone ( PROMPT ) in Cardiac MRI. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70149
BibTeX
@article{Pan2026,
author = {Pan, Yue and Jin, Ning and Chow, Kelvin and Bacher, Mario and Speier, Peter and Yildiz, Vedat O and Tong, Matthew and Han, Yuchi and Ahmad, Rizwan and Simonetti, Orlando P},
title = {Prospective Respiratory Motion Correction Using Machine Learning and Pilot Tone ( PROMPT ) in Cardiac MRI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70149}
}
EUVIMED