Magnitude preparation‐based MR‐acoustic radiation force imaging

Anuj Sharma iD, Kristen Zarcone iD, William A Grissom iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30562 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Anuj Sharma und 2 weiteren Autor:innen. Beteiligte Einrichtung: Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Elasticity Imaging Techniques, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIMH NIH HHS, NINDS NIH HHS, NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To visualize ultrasonic foci in tissue without phase subtraction, to reduce sensitivity to motion and respiration. A secondary purpose is to provide inherently fused anatomy and focus depiction across various image contrasts.

Methods

A magnitude‐contrast MR‐acoustic radiation force imaging (Mag‐ARFI) magnetization‐preparation sequence was implemented that converts displacement‐induced phase shifts into modulated longitudinal magnetization that can be read out by a subsequent imaging sequence. Phantom experiments compared focus visualization between conventional phase contrast‐ARFI and Mag‐ARFI, and demonstrated the ability to quantify displacement using multiple Mag‐ARFI measurements. In vivo phase contrast‐ARFI maps and ‐ and ‐weighted Mag‐ARFI images were acquired using a loop coil that was pulsed with DC to mimic an ultrasound focus‐induced phase shift. Breathing motion artifacts in head images were compared between the phase‐contrast MR‐ARFI and Mag‐ARFI methods.

Results

Phantom experiments demonstrated strong agreement (R = 0.88) between Mag‐ARFI and phase contrast‐ARFI displacement values. Mag‐ARFI and phase‐contrast ARFI peak displacements were 3.02 m and 3.91 m, respectively. Mag‐ARFI FWHM was [9.45, 10.35] mm, compared to phase contrast‐ARFI FWHM of [7.65, 8.56] mm. Head Mag‐ARFI images showed the focus fused with background anatomy, correlating well with phase contrast‐ARFI (R = 0.87 for T 1 ‐weighted, R = 0.79 for T 2 ‐weighted). Mag‐ARFI had reduced breathing motion artifacts, with 3.5 lower coefficient of variation than phase contrast‐ARFI.

Conclusion

Mag‐ARFI allows ultrasound focus localization, with reduced breathing motion artifacts, on ‐ and ‐weighted magnitude images containing background anatomical details.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingElasticity Imaging TechniquesImage Processing, Computer-AssistedPhantoms, ImagingHumansArtifactsBrainRespirationHeadAlgorithmsMotionMaleAdult

Stichwörter der Autor:innen

MR‐ARFIfocused ultrasound (FUS)transcranial ultrasound stimulation (TUS)

Autor:innen und Institutionen

  • Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.

ORCID-Profile: Anuj Sharma, Kristen Zarcone, William A Grissom

Förderung

  • NIMH NIH HHS
  • NINDS NIH HHS
  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Sharma, A., Zarcone, K., & Grissom, W. A. (2025). Magnitude preparation‐based MR‐acoustic radiation force imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30562

Vancouver

Sharma A, Zarcone K, Grissom WA. Magnitude preparation‐based MR‐acoustic radiation force imaging. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30562

BibTeX
@article{Sharma2025,
  author = {Sharma, Anuj and Zarcone, Kristen and Grissom, William A},
  title = {Magnitude preparation‐based MR‐acoustic radiation force imaging},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30562}
}
Kostenlos starten →