DiFiR‐CT: Distance field representation to resolve motion artifacts in computed tomography

Kunal Gupta, Brendan Colvert iD, Zhennong Chen iD, Francisco Contijoch iD

Medical Physics · 2023

DOI: 10.1002/mp.16157 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2023. Verfasst von Kunal Gupta und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Computer Science Engineering, University of California San Diego, San Diego, California, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Movement. Die Arbeit nennt 63 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NHLBI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Motion during data acquisition leads to artifacts in computed tomography (CT) reconstructions. In cases such as cardiac imaging, not only is motion unavoidable, but evaluating the motion of the object is of clinical interest. Reducing motion artifacts has typically been achieved by developing systems with faster gantry rotation or via algorithms which measure and/or estimate the displacement. However, these approaches have had limited success due to both physical constraints as well as the challenge of estimating non‐rigid, temporally varying, and patient‐specific motion fields.

Purpose

To develop a novel reconstruction method which generates time‐resolved, artifact‐free images without estimation or explicit modeling of the motion.

Methods

We describe an analysis‐by‐synthesis approach which progressively regresses a solution consistent with the acquired sinogram. In our method, we focus on the movement of object boundaries. Not only are the boundaries the source of image artifacts, but object boundaries can simultaneously be used to represent both the object as well as its motion over time without the need for an explicit motion model. We represent the object boundaries via a signed distance function (SDF) which can be efficiently modeled using neural networks. As a result, optimization can be performed under spatial and temporal smoothness constraints without the need for explicit motion estimation.

Results

We illustrate the utility of DiFiR‐CT in three imaging scenarios with increasing motion complexity: translation of a small circle, heart‐like change in an ellipse's diameter, and a complex topological deformation. Compared to filtered backprojection, DiFiR‐CT provides high quality image reconstruction for all three motions without hyperparameter tuning or change to the architecture. We also evaluate DiFiR‐CT's robustness to noise in the acquired sinogram and found its reconstruction to be accurate across a wide range of noise levels. Lastly, we demonstrate how the approach could be used for multi‐intensity scenes and illustrate the importance of the initial segmentation providing a realistic initialization. Code and supplemental movies are available at https://kunalmgupta.github.io/projects/DiFiR‐CT.html .

Conclusions

Projection data can be used to accurately estimate a temporally‐evolving scene without the need for explicit motion estimation using a neural implicit representation and analysis‐by‐synthesis approach.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedMovementHumansMotionAlgorithmsHeartArtifactsRotationImage Processing, Computer-AssistedPhantoms, Imaging

Stichwörter der Autor:innen

computed tomographymotion artifactsneural implicit representation

Autor:innen und Institutionen

  • Department of Computer Science Engineering, University of California San Diego, San Diego, California, USA.
  • Shu Chien-Gene Lay Department of Bioengineering, University of California San Diego, La Jolla, California, USA.
  • Department of Radiology, University of California San Diego, La Jolla, California, USA.

ORCID-Profile: Brendan Colvert, Zhennong Chen, Francisco Contijoch

Förderung

  • NHLBI NIH HHS

Zitieren

APA

Gupta, K., Colvert, B., Chen, Z., & Contijoch, F. (2023). DiFiR‐CT: Distance field representation to resolve motion artifacts in computed tomography. Medical Physics. https://doi.org/10.1002/mp.16157

Vancouver

Gupta K, Colvert B, Chen Z, Contijoch F. DiFiR‐CT: Distance field representation to resolve motion artifacts in computed tomography. Medical Physics. 2023. doi:10.1002/mp.16157

BibTeX
@article{Gupta2023,
  author = {Gupta, Kunal and Colvert, Brendan and Chen, Zhennong and Contijoch, Francisco},
  title = {DiFiR‐CT: Distance field representation to resolve motion artifacts in computed tomography},
  journal = {Medical Physics},
  year = {2023},
  doi = {10.1002/mp.16157}
}
Kostenlos starten →