EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Highly Reproducible, Vendor‐Agnostic, Motion‐Insensitive Liver PDFF Mapping at 0.55T , 1.5T , and 3T

Jiayi Tang iD, Daiki Tamada iD, Jon-Fredrik Nielsen iD, Jitka Starekova iD, Julius F Heidenreich iD, Felix Schön iD, Alexandra A Anagnostopoulos iD, Amirhossein Roshanshad iD, Lu Mao iD, Shohei Fujita iD, Pengcheng Xu iD, Christopher Keen iD und 16 weitere

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.70223 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Jiayi Tang und 27 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Liver, Image Processing, Computer-Assisted, Adipose Tissue. Die Arbeit nennt 78 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, Department of Radiology, University of Wisconsin-Madison, NINDS NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop and validate a vendor‐agnostic, motion‐insensitive proton‐density fat‐fraction (PDFF) quantification method.

Methods

Flip‐angle‐modulated (FAM) 2D chemical‐shift‐encoded (CSE) MRI for PDFF quantification was implemented in both the vendor‐agnostic platform Pulseq (“Pulseq‐FAM”) and one vendor‐specific platform (“GE‐specific FAM”). These implementations were distributed to four sites with twelve MR systems of three vendors (Siemens/GE/Philips) and field strengths (0.55T/1.5T/3T). A sequentially‐shipped 16‐vial phantom (PDFF = 0%–30%/T1 water = 200–1400 ms) underwent confounder‐corrected PDFF mapping with commercial 3D‐CSE methods and GE‐specific FAM as available on each system, and Pulseq‐FAM on every system. To assess bias, phantom PDFF measurements were compared to reference. Between‐system variance was evaluated with linear mixed‐effects modeling. Different volunteers were also imaged at each site to assess free‐breathing PDFF mapping feasibility. A prospective single‐site volunteer study was also conducted. Adult patients and children were imaged with breath‐held 3D‐CSE and free‐breathing GE‐specific and Pulseq‐FAM. Radiologists evaluated images for overall quality and motion artifacts. To assess bias, Pulseq‐FAM PDFF measurements were compared to 3D‐CSE and GE‐specific FAM. Test–retest repeatability was assessed by re‐imaging after repositioning. Between‐field‐strength reproducibility was assessed at 1.5T and 3.0T.

Results

In the multi‐center study, Pulseq‐FAM showed reduced T1‐bias and between‐system variability versus 3D‐CSE in phantom PDFF measurements, and free‐breathing feasibility in volunteers. In the single‐site volunteer study ( N = 57), Pulseq‐FAM improved image quality and motion artifacts versus 3D‐CSE ( p < 0.01). Pulseq‐FAM showed excellent agreement with 3D‐CSE (95% limits‐of‐agreement (LoA) = 3.4% PDFF) and GE‐specific FAM (LoA = 2.0%). Pulseq‐FAM showed excellent repeatability (repeatability coefficient (RC) = 1.6% PDFF) and between‐field‐strength reproducibility (reproducibility coefficient (RDC) = 2.4%) versus 3D‐CSE (RC = 2.7%/RDC = 3.4%; differences p < 0.05).

Conclusion

Pulseq‐FAM enables accurate, reproducible, vendor‐agnostic, and motion‐insensitive PDFF quantification in adults and children.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingLiverImage Processing, Computer-AssistedAdipose TissueHumansPhantoms, ImagingReproducibility of ResultsAdultMaleFemaleMotionProspective StudiesImaging, Three-DimensionalArtifactsMiddle Aged

Stichwörter der Autor:innen

MASLDPDFFPulseqcross‐vendorfat quantificationliverproton density fat fractionvendor‐agnostic

Autor:innen und Institutionen

  • Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Functional MRI Laboratory, Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
  • Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany.
  • Institute and Polyclinic for Diagnostic and Interventional Radiology, Faculty of Medicine and University Hospital Carl Gustav Carus Dresden, Dresden University of Technology, Dresden, Germany.
  • Biostatistics and Medical Informatics, University of Wisconsin-Madison, Madison, Wisconsin, USA.

ORCID-Profile: Jiayi Tang, Daiki Tamada, Jon-Fredrik Nielsen, Jitka Starekova, Julius F Heidenreich, Felix Schön, Alexandra A Anagnostopoulos, Amirhossein Roshanshad, Lu Mao, Shohei Fujita, Pengcheng Xu, Christopher Keen, Imam Ahmed Shaik, Eugene Milshteyn, Seonghwan Yee, Andrew J Ellison, David Rutkowski, Jeff Kammerman, Jean H Brittain, Xiaodong Zhong, William A Grissom, Maxim Zaitsev, Aaron L Carrel, Yogesh Rathi, Yun Jiang, Berkin Bilgic, Scott B Reeder, Diego Hernando

Förderung

  • NIBIB NIH HHS
  • Department of Radiology, University of Wisconsin-Madison
  • NINDS NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering

Zitieren

APA

Tang, J., Tamada, D., Nielsen, J. F., Starekova, J., Heidenreich, J. F., Schön, F., Anagnostopoulos, A. A., Roshanshad, A., Mao, L., Fujita, S., Xu, P., Keen, C., Shaik, I. A., Milshteyn, E., Yee, S., Ellison, A. J., Rutkowski, D., Kammerman, J., Brittain, J. H., …, & Hernando, D. (2025). Highly Reproducible, Vendor‐Agnostic, Motion‐Insensitive Liver PDFF Mapping at 0.55T , 1.5T , and 3T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70223

Vancouver

Tang J, Tamada D, Nielsen JF, Starekova J, Heidenreich JF, Schön F, et al. Highly Reproducible, Vendor‐Agnostic, Motion‐Insensitive Liver PDFF Mapping at 0.55T , 1.5T , and 3T. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.70223

BibTeX
@article{Tang2025,
  author = {Tang, Jiayi and Tamada, Daiki and Nielsen, Jon-Fredrik and Starekova, Jitka and Heidenreich, Julius F and Schön, Felix and Anagnostopoulos, Alexandra A and Roshanshad, Amirhossein and Mao, Lu and Fujita, Shohei and Xu, Pengcheng and Keen, Christopher and Shaik, Imam Ahmed and Milshteyn, Eugene and Yee, Seonghwan and Ellison, Andrew J and Rutkowski, David and Kammerman, Jeff and Brittain, Jean H and Zhong, Xiaodong and Grissom, William A and Zaitsev, Maxim and Carrel, Aaron L and Rathi, Yogesh and Jiang, Yun and Bilgic, Berkin and Reeder, Scott B and Hernando, Diego},
  title = {Highly Reproducible, Vendor‐Agnostic, Motion‐Insensitive Liver PDFF Mapping at 0.55T , 1.5T , and 3T},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.70223}
}
Kostenlos starten →