EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Novel 1 H/ 19 F double‐tuned coil using an asymmetrical butterfly coil

Suk-Min Hong, Chang-Hoon Choi, Jörg Felder, N Jon Shah

Medical Physics · 2025

DOI: 10.1002/mp.17890 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Suk-Min Hong und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Neuroscience and Medicine 4, INM-4, Forschungszentrum Jülich, Jülich, Germany. Hauptthemen laut MEDLINE: Protons, Magnetic Resonance Imaging, Fluorine. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft, Land Nordrhein-Westfalen (NRW).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Fluorine‐19 ( 19 F) magnetic resonance imaging (MRI) is a non‐invasive imaging tool for the targeted application of fluorinated agents, such as cell tracking, and for the demonstration of oximetry. However, as the SNR of 19 F is significantly weaker than that of proton ( 1 H) imaging, the 19 F coil must be combined with 1 H coils for anatomical co‐registration and B 0 shimming. This is difficult due to the strong coupling between the coils when they are in proximity, and is problematic since the Larmor frequency of 19 F is 94% that of 1 H, further increasing the potential for coupling between the 1 H and 19 F elements.

Purpose

Conventional double‐tuned coil methods tend to generate loss compared to single‐tuned reference coils. The asymmetrical butterfly coil has a split resonance peak, which can cover frequencies of 1 H and 19 F without losses arising from lossy traps or switching circuits. In this study, the use of an asymmetrical butterfly coil was evaluated for 1 H/ 19 F applications.

Methods

To increase quadrature efficiency at both the 1 H and 19 F frequencies, the left and right loops of the butterfly coil were tuned asymmetrically. The coil's tuning and performance were evaluated in simulations and MR measurements, and the results were compared to a dimension‐matched single‐tuned loop coil.

Results

The split resonance peak of the asymmetrical butterfly coil successfully spanned the 19 F to 1 H frequency. It operated with higher quadrature efficiency at both 1 H and 19 F frequencies and demonstrated superior receive sensitivity and SNR compared to the dimension‐matched single‐tuned loop coil.

Conclusions

The split resonance peak of the asymmetrical butterfly coil supported both 1 H and 19 F frequencies, delivering a higher SNR than that of the single‐tuned loop coil. Since the asymmetrical butterfly coil can cover ¹H and ¹⁹F frequencies without loss and provides higher efficiency than the reference single‐tuned coil, it can be effectively utilized for ¹H/¹⁹F MRI applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ProtonsMagnetic Resonance ImagingFluorineEquipment DesignPhantoms, ImagingSignal-To-Noise Ratio

Stichwörter der Autor:innen

19FMRIRF coil

Substanzen

ProtonsFluorine

Autor:innen und Institutionen

  • Institute of Neuroscience and Medicine 4, INM-4, Forschungszentrum Jülich, Jülich, Germany.
  • Institute of Neuroscience and Medicine 11, INM-11, JARA, Forschungszentrum Jülich, Jülich, Germany.
  • JARA - BRAIN - Translational Medicine, Aachen, Germany.
  • Department of Neurology, RWTH Aachen University, Aachen, Germany.

Förderung

  • Deutsche Forschungsgemeinschaft
  • Land Nordrhein-Westfalen (NRW)

Zitieren

APA

Hong, S. M., Choi, C. H., Felder, J., & Shah, N. J. (2025). Novel 1 H/ 19 F double‐tuned coil using an asymmetrical butterfly coil. Medical Physics. https://doi.org/10.1002/mp.17890

Vancouver

Hong SM, Choi CH, Felder J, Shah NJ. Novel 1 H/ 19 F double‐tuned coil using an asymmetrical butterfly coil. Medical Physics. 2025. doi:10.1002/mp.17890

BibTeX
@article{Hong2025,
  author = {Hong, Suk-Min and Choi, Chang-Hoon and Felder, Jörg and Shah, N Jon},
  title = {Novel 1 H/ 19 F double‐tuned coil using an asymmetrical butterfly coil},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17890}
}
Kostenlos starten →