Analytical Dual Flip Angle R1 Calculation Outside the Small‐Angle Regime
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70174 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Luke J Edwards und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurophysics, Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Image Processing, Computer-Assisted. Die Arbeit nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Seventh Framework Programme, Horizon 2020 Framework Programme, Max-Planck-Gesellschaft.
Abstract (Originalsprache)
Purpose
To evaluate a new analytical estimator for R1 and apparent proton density () from short‐TR dual flip angle data which does not rely on the small flip angle approximation and can thus be applied to a broader range of data, especially where relatively large flip angles are needed to achieve sufficient T1‐weighting for R1 estimation. Theory and Methods A rational approximation of the Ernst equation was derived for small and rearranged to give analytical estimators of R1 and from dual flip angle data. Unlike previously used analytical estimators, this method relies neither on the flip angles being small nor the two TRs being equal or integer multiples of each other. R1 and estimated using the novel method were compared to estimates using the conventional small‐angle approximation approach in simulations and data measured at 7T from six in vivo human participants and a postmortem chimpanzee brain. Test–retest scans of the six participants were used to evaluate within‐participant coefficients of variance (WCV) of R1 and using the two methods.
Results
The small‐angle approximation gave rise to a flip angle‐dependent bias in all cases. This bias was not observed using the novel method, demonstrating its higher accuracy. There were only negligible differences in WCV between the methods, demonstrating that precision is preserved.
Conclusion
The increase in accuracy and preservation of precision suggest that the novel method should be used instead of the current small flip angle method.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Neurophysics, Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany.
- Department of Cognitive Neuroscience, Faculty of Psychology and Neuroscience, Maastricht University, Maastricht, the Netherlands.
- Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, Sweden.
- Felix Bloch Institute for Solid State Physics, Faculty of Physics and Earth System Sciences, Leipzig University, Leipzig, Germany.
- Wellcome Centre for Human Neuroimaging, Institute of Neurology, University College London, London, UK.
ORCID-Profile: Luke J Edwards, Kerrin J Pine, Ilona Lipp, Evgeniya Kirilina
Förderung
- Seventh Framework Programme
- Horizon 2020 Framework Programme
- Max-Planck-Gesellschaft
- Federal Ministry of Education and Research (BMBF)
Zitieren
APA
Edwards, L. J., Pine, K. J., Lipp, I., EBC consortium, Kirilina, E., Helms, G., & Weiskopf, N. (2026). Analytical Dual Flip Angle R1 Calculation Outside the Small‐Angle Regime. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70174
Vancouver
Edwards LJ, Pine KJ, Lipp I, EBC consortium, Kirilina E, Helms G, et al. Analytical Dual Flip Angle R1 Calculation Outside the Small‐Angle Regime. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70174
BibTeX
@article{Edwards2026,
author = {Edwards, Luke J and Pine, Kerrin J and Lipp, Ilona and EBC consortium, EBC consortium and Kirilina, Evgeniya and Helms, Gunther and Weiskopf, Nikolaus},
title = {Analytical Dual Flip Angle R1 Calculation Outside the Small‐Angle Regime},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70174}
}