Fast and silent MRI using nonlinear gradient fields at the ultrasonic gradient switching frequency of 20 kHz with a Point Spread Function framework reconstruction
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30230 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Michael J B McGrory und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Center for Image Sciences, University Medical Center Utrecht, Utrecht, The Netherlands. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Algorithms, Phantoms, Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nederlandse Organisatie voor Wetenschappelijk Onderzoek.
Abstract (Originalsprache)
Purpose
To demonstrate the feasibility of using a nonlinear gradient field for spatial encoding at the ultrasonic switching frequency of 20 kHz and present a framework to reconstruct data acquired in this way.
Methods
Nonlinear encoding at 20 kHz was realized by using a single‐axis silent gradient insert for imaging in the periphery, that, is the nonlinear region, of the gradient field. The gradient insert induces a rapidly oscillating gradient field in the phase‐encode direction, which enables nonlinear encoding when combined with a Cartesian readout from the linear whole‐body gradients. Data from a 2D gradient echo sequence were reconstructed using a point spread function (PSF) framework. Accelerated scans were also simulated via retrospective undersampling ( R = 1 to R = 8) to determine the effectiveness of the PSF‐framework for accelerated imaging.
Results
Using a nonlinear gradient field switched at 20 kHz and the PSF‐framework resulted in images of comparable quality to images from conventional Cartesian linear encoding. At increased acceleration factors ( R ≤ 8), the PSF‐framework outperformed linear SENSE reconstructions by improved controlling of aliasing artifacts.
Conclusion
Using the PSF‐framework, images of comparable quality to conventional SENSE reconstructions are possible via combining traditional linear and ultrasonic oscillating nonlinear encoding fields. Using nonlinear gradient fields relaxes the demand for strictly linear gradient fields, enabling much higher slew rates with a reduced risk of peripheral nerve stimulation or cardiac stimulation, which could aid in extension to ultrasonic whole‐body MRI. The lack of aliasing artifacts also highlights the potential of accelerated imaging using the PSF‐framework.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Center for Image Sciences, University Medical Center Utrecht, Utrecht, The Netherlands.
- Spinoza Center for Neuroimaging Amsterdam, Amsterdam, The Netherlands.
ORCID-Profile: Michael J B McGrory, Edwin Versteeg, Jeroen C W Siero
Förderung
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek
Zitieren
APA
McGrory, M. J. B., Versteeg, E., Sbrizzi, A., van den Berg, C. A. T., Klomp, D., & Siero, J. C. W. (2024). Fast and silent MRI using nonlinear gradient fields at the ultrasonic gradient switching frequency of 20 kHz with a Point Spread Function framework reconstruction. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30230
Vancouver
McGrory MJB, Versteeg E, Sbrizzi A, van den Berg CAT, Klomp D, Siero JCW. Fast and silent MRI using nonlinear gradient fields at the ultrasonic gradient switching frequency of 20 kHz with a Point Spread Function framework reconstruction. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30230
BibTeX
@article{McGrory2024,
author = {McGrory, Michael J B and Versteeg, Edwin and Sbrizzi, Alessandro and van den Berg, Cornelis A T and Klomp, Dennis and Siero, Jeroen C W},
title = {Fast and silent MRI using nonlinear gradient fields at the ultrasonic gradient switching frequency of 20 kHz with a Point Spread Function framework reconstruction},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30230}
}
EUVIMED