Free‐breathing 3D pulmonary ventilation mapping at 0. 55 T using stack‐of‐spiral out‐in bSSFP
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70069 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2026. Written by Ziwei Zhao and 7 co-authors. Institution: Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, Californ…. Major MEDLINE topics: Imaging, Three-Dimensional, Lung, Magnetic Resonance Imaging, Pulmonary Ventilation. Cited 2 times so far (Crossref); 59 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: National Science Foundation, National Science Foundation of Sri Lanka.
Abstract (original language)
Purpose
To develop a free‐breathing pulmonary imaging technique that provides three‐dimensional ( 3D) structural images and regional ventilation maps, and to evaluate its repeatability and accuracy compared with two‐dimensional phase‐resolved functional lung ( PREFUL ) and global tidal volume.
Methods
A free‐breathing 3D stack‐of‐spiral out‐in ( SOS out‐in) balanced steady‐state free precession (bSSFP) sequence with self‐navigators was designed to achieve 2‐mm isotropic resolution in 5 min. Respiratory‐resolved images were reconstructed using spatial L1 wavelet and temporal finite‐difference constraints. The 3D ventilation maps were generated based on Jacobian determinant of the estimated nonrigid deformations. Six healthy volunteers were scanned in supine and prone positions. Ventilation maps were compared with two‐dimensional PREFUL from two matched slices. Test–retest repeatability was assessed using Bland–Altman analysis. Correlations among the proposed method, PREFUL, and global tidal volume were evaluated.
Results
In healthy volunteers, the SOS out‐in lung images provided sufficient vessel‐parenchyma contrast and boundary sharpness to support accurate ventilation estimation. Regional ventilation measurements from 3D SOS out‐in demonstrated good repeatability (relative differences < 10%). Ventilation maps from 3D SOS out‐in strongly correlated with PREFUL on a slice‐matched basis as well as with global tidal volume ( R 2 > 0.7, p < 0.001).
Conclusion
The proposed method provides high‐quality respiratory‐resolved structural images and 3D ventilation mapping in a single 5‐min scan at 0. 55 T . Ventilation measurements are sensitive, consistent, and in good agreement with PREFUL and spirometry.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, California, USA.
- Vista AI, Inc., Palo Alto, California, USA.
- Alfred E. Mann Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, California, USA.
- Department of Radiology, Children's Hospital Los Angeles, Los Angeles, California, USA.
- Siemens Medical Solutions USA, Inc, Los Angeles, California, USA.
ORCID profiles: Ziwei Zhao, Nam G Lee, Bilal Tasdelen, Ye Tian, Bochao Li, Sophia X Cui, Krishna S Nayak
Funding
- National Science Foundation
- National Science Foundation of Sri Lanka
Cite
APA
Zhao, Z., Lee, N. G., Tasdelen, B., Miao, X., Tian, Y., Li, B., Cui, S. X., & Nayak, K. S. (2026). Free‐breathing 3D pulmonary ventilation mapping at 0. 55 T using stack‐of‐spiral out‐in bSSFP. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70069
Vancouver
Zhao Z, Lee NG, Tasdelen B, Miao X, Tian Y, Li B, et al. Free‐breathing 3D pulmonary ventilation mapping at 0. 55 T using stack‐of‐spiral out‐in bSSFP. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70069
BibTeX
@article{Zhao2026,
author = {Zhao, Ziwei and Lee, Nam G and Tasdelen, Bilal and Miao, Xin and Tian, Ye and Li, Bochao and Cui, Sophia X and Nayak, Krishna S},
title = {Free‐breathing 3D pulmonary ventilation mapping at 0. 55 T using stack‐of‐spiral out‐in bSSFP},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70069}
}
EUVIMED